RIGOL
VSA Programming Guide
2-49
:SENSe:DDEMod:FILTer:REFerence GAUSsian
The following query returns GAUS.
:SENSe:DDEMod:FILTer:REFerence?
[:SENSe]:DDEMod:FILTer:REFerence:USER<n>[:DATA]
Syntax
[:SENSe]:DDEMod:FILTer:REFerence:USER<n>[:DATA] <ampl>{,<ampl>}
[:SENSe]:DDEMod:FILTer:REFerence:USER<n>[:DATA]?
Description
Sets the data of the specified reference user-defined filter.
Queries the data of the specified reference user-defined filter.
Parameter
Name
Type
Range
Default
<n>
Discrete
1|2|3|4|5|6
——
<ampl>
Consecutive Real Number
——
——
Return Format
The query returns the user-defined filte data in scientific notation.
Example
The following command sets the data of the reference user-defined filter 1.
:SENSe:DDEMod:FILTer:REFerence:USER1:DATA
-0.000321525,0.00294817,0.002500879,-0.001910775,-0.005052538
The following query returns
-3.215250000e-04,2.948170000e-03,2.500879000e-03,-1.910775000-03,-5.052538000e-03
。
:SENSe:DDEMod:FILTer:REFerence:USER1:DATA?
[:SENSe]:DDEMod:FSK:DEViation:REFerence
Syntax
[:SENSe]:DDEMod:FSK:DEViation:REFerence <freq>
[:SENSe]:DDEMod:FSK:DEViation:REFerence?
Description
Sets the reference frequency offset in the FSK modulation.
Queries the reference frequency offset in the FSK modulation.
Parameter
Name
Type
Range
Default
<freq>
Consecutive Real Number
1 kHz to 1T Hz
1 kHz
Remarks
This command is only valid when you select FSK as the modulation format and reference deviation mode is
set to Manual.
Return Format
The query returns the reference frequency offset in scientific notation. The unit is Hz.