Chapter 2 Functions of the Front Panel of GPSA
RIGOL
RSA3000E User Guide
2-45
You can use the numeric keys, the knob, or arrow keys to modify this parameter;
you can also use the touch screen to modify the parameter.
Table 2-33 Stop Line for T-Power Measurement
Parameter
Remarks
Default
1 ms
Range
current value of start line to
current value of sweep time
Unit
s, ms, μs, ns, ps
Knob Step
sweep time/600, Min = 1 μs
Left/Right Arrow Key Step
Up/Down Arrow Key Step
at 1-1.5-2-3-5-7.5 step
7.
Auto Couple
When you enable "Auto Couple" function, all the manual/auto settings in the
current measurement mode will be set to "Auto".
In auto state, the auto coupled parameters are changed with their coupled
parameters. The auto coupling operation will ensure the optimal performance of
the instrument. After the operation, all the auto coupled parameters will
immediately be automatically reset based on the coupled parameters.
8.
Meas Preset
Restores all parameters of the current measurement mode to their factory
default values.