RP1002C User’s Guide
Specifications
These characteristics apply to an adjusted RP1002C current probe
installed on an oscilloscope of any brand. The oscilloscope must be
warmed up for at least 20 minutes and be in an environment with
the temperature at 10
℃
-30
℃
and the humidity at 0-80%.
Electrical Characteristics
Current Range
50mV/A, 500mV/A
DC Accuracy, typical
500mV/A: ±3% ±20mA
(20mA to 14A peak range)
50mV/A: ±4% ±200mA
(200mA to 100A peak range)
50mV/A: ±15% maximum
(100A peak to 140A peak range)
Gain versus frequency,
typical
See Appendix 1 Gain versus
Frequency at 1 A Peak (Typical)
Max. Working Current
See
Voltage and Current
Ratings
Max. Working Voltage
See
Voltage and Current
Ratings
Max. Floating Voltage
See
Voltage and Current
Ratings
Frequency Range
DC to 1MHz (-3dB)
Battery Type and Life,
typical
9V NEDA 1604A, IEC 6LR61
8 hours (1 each)
DC Signal Linearity,
typical
See Appendix 3 DC Signal
Linearity in 50mV/A Range
(Typical)
Phase versus
frequency, typical
See Appendix 4 Phase versus
Frequency at 1 A Peak (Typical)