![Rigol DS4000E Series Скачать руководство пользователя страница 202](http://html1.mh-extra.com/html/rigol/ds4000e-series/ds4000e-series_programming-manual_1460996202.webp)
RIGOL
2-190
DS4000E Programming Guide
DelayRange = 14 x DelayScale.
MainScale is the main time base, refer to the :TIMebase[:MAIN]:SCALe command; MainOffset is the
horizontal position (normal sweep), refer to the :TIMebase[:MAIN]:OFFSet command; DelayScale is the
horizontal time base of the delayed sweep, refer to the :TIMebase:DELay:SCALe command.
Return Format
The query returns the horizontal position of the delayed sweep in scientific notation.
:TIMebase:DELay:SCALe
Syntax
:TIMebase:DELay:SCALe <scale>
:TIMebase:DELay:SCALe?
Description
Sets or queries the horizontal time base of the delayed sweep. The default unit is s/div.
Parameter
Name
Type
Range
Default
<scale>
Real
The minimum value of <scale> is related to the current
sample rate. It must satisfy the formula (1x50/real-time
sample rate x 1/40) and should not be less than 2 ns. The
maximum value of <scale> is the current main time
base.
500 ns/div
Return Format
The query returns the horizontal time base of the delayed sweep in scientific notation.
:TIMebase:HREF:MODE
Syntax
:TIMebase:HREF:MODE <href>
:TIMebase:HREF:MODE?
Description
Sets or queries the horizontal reference.
Parameter
Name
Type
Range
Default
<href>
Discrete
{CENTer|TPOSition|USER}
CENTer
Remarks
CENTer: when the horizontal time base is modified, the waveform displayed will be expanded or
compressed horizontally relative to the screen center.
TPOSition: when the horizontal time base is modified, the waveform displayed will be expanded or
compressed horizontally relative to the trigger position.
USER: when the horizontal time base is modified, the waveform displayed will be expanded or
compressed horizontally relative to the user-defined reference position (refer to
the :TIMebase:HREF:POSition command).
Only when the horizontal time base mode is Y-T (refer to the :TIMebase:MODE command) and the
delayed sweep is off (refer to the :TIMebase:DELay:ENABle command), can this command be valid.
Return Format
The query returns CENT, TPOS, or USER.
Содержание DS4000E Series
Страница 1: ...RIGOL Programming Guide DS4000E Series Digital Oscilloscope Aug 2016 RIGOL TECHNOLOGIES INC...
Страница 2: ......
Страница 266: ...RIGOL Chapter 3 Programming Examples 3 6 DS4000E Programming Guide...
Страница 270: ...RIGOL Chapter 3 Programming Examples 3 10 DS4000E Programming Guide...