Chapter 2 Constant Voltage Tests
DP1308A Performance Verification Manual
2-2
Preparation
The main parameters of DP1308A requiring test under constant voltage include
constant voltage load adjustment rate, constant voltage linear adjustment rate, ripple
and noise, transient response time as well as programming and readback accuracy.
Before the test, select the appropriate voltage (for the voltages supported by
DP1308A, refer to Table 2-1) via the “voltage selector” at the rear panel of the power
supply according to the AC line voltage of your nation. Connect the device
recommended in Recommended Test Devices according to the figure below under
normal temperature (about 25
℃
). During the test, set the voltage of the AC source
according to the voltage selected.
DP1308A
+
-
Electronic
Load
+
-
AC Source
Multimeter/
Oscilloscope/
RMS Voltmeter
-
+
+
-
+
-
Figure 2-1 Test Connection (CV)
Table 2-1
AC line voltages supported by DP1308A
Voltage
100VAC
115VAC
220VAC
230VAC
Undervoltage 90VAC
104VAC
242VAC
253VAC
Overvoltage
110VAC
127VAC
198VAC
207VAC