Troubleshooting and Customer Service

For more information on error messages and trouble shooting, please refer to the Error Messages and 
Trouble Shooting section of RIGHTEST User's Manual.
If you have any questions or in case of problems with RIGHTEST products, please contact local Bionime 
distributor or email to [email protected] .

Lay User Evaluation 

A total of 110 layperson were enrolled. Each patient tested their  fingertip blood samples with 3 lots of Wiz 
blood glucose test strip and Wiz Plus meter. Then the professional collected blood samples were 
centrifuged immediately after collection to obtain plasma. Analyze the plasma by the lab instrument (YSI 
2300 analyzer). 100% of the RIGHTEST Wiz Plus BGMS were within ± 15% of YSI values at glucose 
concentrations 

 100 mg/dL (5.55 mmol/L) and within ± 15 mg/dL (0.83 mmol/L) at glucose 

concentrations < 100 mg/dL (5.55 mmol/L).

Hematocrit

Hematocrit should be between 20 to 60%. If you do not know your hematocrit, ask your healthcare 
professional.

Interferences

26 toxic amount tested substances (Acetaminophen, Ascorbic acid, Dopamine, EDTA, Gentisic acid, 
Heparin, Ibuprofen, L-dopa, Methyldopa, Pralidoxime iodide, Salicylic Acid, Tetracycline, Tolazamide,
Tolbutamide, Bilirubin, Cholesterol, Creatinine, Glutathione, Haemoglobin, Triglycerides, Uric acid, Maltose, 
Xylose, Galactose, Lactose, Icodextrin) in two blood sample concentrations.
Substance and possible level may interfere the glucose measurement:
Ascorbic acid 

 6 mg/dL (0.34 mmol/L)  

Glutathione reduced 

 70 mg/dL (2.28 mmol/L)

Uric Acid 

 16 mg/dL (0.95 mmol/L)

Reagents

Each Blood Glucose Test Strip contains the following reagents:
Glucose Oxidase (GOD)                  18.8 %          
Potassium Ferricyanide                    37.7 %

 

 

 

               

Non-reactive Ingredients                  43.5 %

 

 

 

             

References

1) Diabetes Information - American Association for Clinical Chemistry (AACC)

Electronic Version

    Retrieved May. 08, 2019 from www.labtestsonline.org/understanding/analytes/glucose/test.html
2) Review of Glucose Oxidases and Glucose Dehydrogenases: A Bird's Eye View of Glucose Sensing 
    Enzymes. J Diabetes Sci Technol 2011 Sep; 5(5): 1068–1076 (2011).

Accuracy

The accuracy of 

Blood Glucose Monitoring System was tested by comparing 

RIGHTEST

 Wiz Plus 

fingertip whole blood (plasma equivalent) glucose values measured by 

Blood 

RIGHTEST

 Wiz Plus 

Glucose Meter with plasma glucose values obtained from a YSI 2300 reference instrument.
The YSI 2300 was calibrated with NIST (SRM) 917c reference.
The results and variations between the two methods, 

 

Blood Glucose Monitoring 

RIGHTEST

Wiz Plus 

System and YSI 2300 (as the reference method) are shown in the tables below.

Additional Information for Healthcare Professionals

(ii) Control solution: 

P-01

300

38.6 (2.1)
1.1 (0.06)

2.8 %

P-02

300

84.7 (4.7)
1.7 (0.09)

2.0 %

P-03

300

136.5 (7.6)

2.5 (0.14)

1.8 %

(i) Venous whole blood sample:

Glucose levels
(1) Total test numbers (n)
(2) Mean mg/dL (mmol/L)
(3) SD mg/dL (mmol/L)
(4) CV (%)

Glucose levels
(1) Total test numbers (n)
(2) Mean mg/dL (mmol/L)
(3) SD mg/dL (mmol/L)
(4) CV (%)

(2)

Detection Principle

The glucose oxidase and potassium ferricyanide in the strip react with the glucose in the sample to 
produce an electrical current which is proportional to the amount of glucose in the sample. The meter 
measures the current and converts it to the corresponding glucose concentration.

Performance Characteristics

Data generated using RIGHTEST Wiz Plus Blood Glucose Meter. RIGHTEST Wiz Plus Blood Glucose 
Meter is the representative of the RIGHTEST Wiz Blood Glucose Meter family (included Wiz and Wiz 
Plus Blood Glucose Meter).

Precision

The precision was evaluated including (i) 5 glucose levels of venous whole blood samples (blood 
samples were collected with heparin tube and altered for test after 8 hours) (ii) 3 glucose levels of 
control solution, in period of 10 days, by 10 meters and 3 batches of strips.

P-04

300

221.2 (12.3)

2.9 (0.16)

1.3 %

P-05

300

329.9 (18.3)

4.4 (0.24)

1.3 %

CS-L

300

37.9 (2.1)
1.0 (0.06)

2.6 %

CS-N

300

85.4 (4.7)
1.5 (0.08)

1.8 %

CS-H

300

246.4 (13.7)

2.6 (0.14)

1.0 %

Quality Control Section

Please refer to the Quality Control section of the User's Manual.

Storage and Handling

- Store the strips in the original capped vial at temperatures between 4°C to 30°C (39°F to 86°F) and relative 

humidity below 90 %. Do not freeze.

- Close the vial cap immediately and tightly after taking test strip out from the vial. Do not leave the cap of 

vial opened. If the strip is exposed to the air too long, it will absorb the moisture and cause wrong test 
result.

- When you open a new vial of test strips please write the first opening date on the label. Use test strips 

within 12 months after first opened or until the expiration date printed on the label (whichever comes first).

Measurement Range

The measurement range of RIGHTEST System is 10 to 600 mg/dL or 0.6 to 33.3 mmol/L.

NOTE

- Suggest not to use this meter close to source of strong electromagnetic radiation, to avoid 

interference with proper operation.

- Suggest to keep meter free of dust, water or any liquid.

- Flouoride should not be used as a preservative when collecting blood glucose samples.
- Hands and fingers contaminated with sugar from foods or beverages may cause false elevated results.
- The results of blood glucose measurements are different for measurements with whole blood and 

plasma.

- Storage of strips near bleach as well as bleach containing products will affect the results of the  

RIGHTEST Test Strips.

- RIGHTEST Blood Glucose Test Strips are designed for use with capillary whole blood samples. 
  Do not use serum or plasma samples.
- Incorrect test results may be obtained at high altitude more than about 3,048 meters (10,000 feet) 
  above sea level.
- Severe dehydration and excessive water loss may cause inaccurately low results.
- Patients going through oxygen therapy may yield falsely low results.
- RIGHTEST Blood Glucose Monitoring System has not been validated for use on neonates. 
  Therefore, it should not be used for neonates.
- Do not perform the blood glucose test at temperatures below 10°C (50°F) or above 40°C (104°F), below 

10% or above 90% relative humidity. The suggested temperature range for the control solution test is 
15 to 40°C (59 to 104°F).

Difference range in values
between the YSI value and 
RIGHTEST Wiz Plus meter

The percent ( and number ) of samples was the difference between RIGHTEST 
Wiz Plus blood glucose meter and the YSI value within the following intervals.

Difference range in values
between the YSI value and 
RIGHTEST Wiz Plus meter

Within ± 5 %

Within ± 10 %

Within ± 15 %

The percent ( and number ) of samples was the difference between RIGHTEST 
Wiz Plus blood glucose meter and the YSI value within the following intervals.

*Acceptance criteria in ISO15197 : 2013 are that 95 % of all differences in glucose values should be within 

±

 15 mg/dL (0.83 mmol/L) at glucose concentrations < 100 mg/dL (5.55 mmol/L), and within ± 15 % at 

glucose concentrations 

 100 mg/dL (5.55 mmol/L).

Note: For glucose concentrations < 100 mg/dL (5.55 mmol/L), difference values are expressed in mg/dL 
(mmol/L), and for glucose concentrations 

 100 mg/dL (5.55 mmol/L), difference values are compared in 

percentage.

Fingertip

65.0% (277/426)

98.4% (419/426)

100%(426/426)

Palm

62.4%(266/426)

95.5%(407/426)

100%(426/426)

Forearm

59.6%(254/426)

95.1%(405/426)

100%(426/426)

Fingertip

68.2%(135/198)

97.5%(193/198)

100% (198/198)

Palm

68.2%(135/198)

98.5%(195/198)

100% (198/198)

Forearm

53.5%(106/198)

92.4%(183/198)

100% (198/198)

Within ± 5 mg

/

dL (0.28 mmol/L)

Within ± 10 mg

/

dL (0.56 mmol/L)

Within ± 15 mg

/

dL (0.83 mmol/L)

Table 2: Represents samples for glucose results 

 100 mg/dL (5.55 mmol/L).

Table 1: Represents samples for glucose results < 100 mg/dL (5.55 mmol/L).

Rev. Date:2019-05

For 

in vitro

 diagnostic use

Manufacturer

IVD

For single use only

EC Representive

LOT

Lot number

CE-mark (with No. of notified body)

Consult the instruction for use

Store  between  tempe rature  4°C  and  30°C  ( 39°F and  86 °F) 

 

Biological risks

Expiry date