Back-to-back distance measuring gauge, IMR
IMR [Revision 1.0] 5th November 2010
7
Figure 2
12.2. Get into operational modes
•
To get into the operational modes it is necessary to turn-off the device (press
Red
button more then 3 seconds).
•
Keeping
Green
button in pushed position turn-on the device (press
Red
but-
ton).
•
Display shows “
CLbr.0
” message (the mode of calibration of the sensor's "
0
").
•
To get into this mode it is necessary to press
Green
button (see. p. 11.2).
•
For transition to next mode it is necessary to press
Blue
button.
•
Display shows “
CLbr.b
” message (the mode of device base calibration).
•
To get into this mode press
Green
button (see. p.11.3).
•
To get out the mode of calibration press
Red
button.
12.3. Calibration of the sensor zero
•
Zero calibration mode being enabled, the display shows the length of calibra-
tion tool.
•
If editing of the length value is not required, go to the next step. To edit the
length value, press the
Blue
button, and the digit to be edited starts blinking.
Changing over between the digits is made by pressing the
Blue
button while
changing of values is made by pressing the
Green
button. When editing is fin-
ished, press the
Red
button and confirm or cancel saving of the parameter by
pressing the
Green
button or
Red
button, respectively.
•
Press the
Green
button, and the length value starts blinking, which means
that the device must be placed onto calibration tool. The measurement tip and
support ball of the gauge must be tightly fitted to the flat plates of calibration
tool (figure 3).
•
Press the
Green
button, and the display shows current reading of the sensor
in its own coordinate system. By moving the device, assure that repeatability
of measurement results is obtained.
Figure 3
Содержание IMR Series