Back-to-back distance measuring gauge, IMR-L
IMR-L [Revision 1.0] 1th October 2012
6
10. Indication parameters setup
10.1. Image brightness set-up
To change the display brightness it is necessary to:
•
turn the power on (press
Red
button);
•
press
Blue
button and keep it pressed for more than three seconds;
•
display will show
“Up”
;
•
press
Blue
button one more time, the image brightness set-up will be dis-
played “
brt
”;
•
choose the necessary brightness value by
Green
button pressings;
•
to save the changed parameters press
Red
button the display shows “
SAUE
”
message, press the
Green
button to confirm saving and
Red
to cancel saving
of the changed parameters;
At brightness choosing one should take into account that increased brightness
enhances power consumption and decreases the period till battery recharge moment.
10.2. Millimeters-Inches display set-up
To change the mode it is necessary to:
•
turn the power on (press
Red
button);
•
press
Blue
button and keep it pressed for more than three seconds;
•
display will show “
Up
”;
•
press
Blue
button again until the either “
SI
” or “
Inch
” message appears on the
display. “
SI
” – measuring results will be shown in mm, “
Inch
” – measuring re-
sults will be shown in inches;
•
choose the necessary value by
Green
button pressings;
•
to save the changed parameters press
Red
button the display shows “
SAUE
”
message, press the
Green
button to confirm saving and
Red
to cancel saving
of the changed parameters
NOTE
: Discreteness of inch indication is two digits after point. The separation
point is indicated at the top side of the screen in the reverse indication regime
11. Service operation modes
This section contains the description of the modes for the check of device effi-
ciency and calibration. As erroneous actions in this mode can lead to invalid measure-
ment results, only specially trained personnel should perform such operations.
11.1. Calibration conditions
Calibration of the device is not necessary in the current work. It is neces-
sary only after producing, repairing and also after checking with negative result.
To perform calibration the following means are necessary:
•
Calibration tool RF260.90.000 fig.2;