14
Attention!
Medical electrical equipment is subject to special precautions regarding elec-
tromagnetic compatibility (EMC).
Portable and mobile radio frequency communication devices can affect medical
electrical equipment. The ME device is for operation in an electromagnetic en-
vironment and intended for professional facilities such as industrial areas and
hospitals.
The user of the device should ensure that it is operated within such an en¬vi-
ronment.
Warning!
The ME device may not be stacked, arranged or used directly next to or with
other devices. When operation is required to be close to or stacked with other
devices, the ME device and the other ME devices must be observed in order to
ensure proper operation within this arrangement. This ME device is intended
for use by medical professionals only. This device may cause radio interference
or interfere with the operation of nearby devices. It may become necessary to
take appropriate corrective measures, such as redirecting or rearranging the ME
device or shield.
The rated ME device does not exhibit any basic performance features in the sense
of IEC 60601-1, which would present an unacceptable risk to patients, operators
or third parties should the power supply fail or malfunction.
Warning!
Portable / stationare RF communications equipment (radios) including acces-
sories, such as antenna cables and external antennas, should not be used in
closer proximity than 30 cm (12 inches) to parts and cables of the Electronic
Stethoscopes ri-sonic PCP-USB / ri-sonic PCP-1 specified by the manufacturer.
Failure to comply may result in a reduction in the device‘s performance features.
Directives and manufacturer‘s declaration - Electromagnetic emissions
The Electronic Stethoscopes ri-sonic PCP-USB / ri-sonic PCP-1 are intended for
use in the electromagnetic environ¬ment specified below. The customer or user
of the Electronic Stethoscopes ri-sonic PCP-USB / ri-sonic PCP-1 should ensure
that it is used in such an environment.