Instruction of use

Electronic Stethoscope

ri-

sonic

 PCP-USB

ri-

sonic

 PCP-1

Содержание ri-sonic PCP-1

Страница 1: ...1 Instruction of use Electronic Stethoscope ri sonic PCP USB ri sonic PCP 1 ...

Страница 2: ...al conditions Intended Environment 1 5 Warnings Attention 2 First application 2 1 Scope of Supply 2 2 Device Function 3 Operation and Function 3 1 Introduction 3 2 Commissioning 3 2 1 Installation ri sonic PCP USB 3 2 2 Installation ri sonic PCP 1 3 23 Auscultation Sessions Examination of the patient 4 Care instructions 4 1 General Information 4 2 Cleaning and Disinfection 5 Technical Data 6 Acces...

Страница 3: ...ents described in these instructions for use are only to be used by suitably trained personnel The perfect and safe functioning of this device is only guaranteed when original parts and accessories from Riester are used 1 2 Safety Symbols Information Symbol Note on symbol Follow the instructions in the operation manual Protection class II IP22 Warning The general warning sign indicates a potential...

Страница 4: ...tfeuchtigkeit für Transport und Lagerbedingungen Air pressure for transport and storage Air pressure for Ambient Operating 0124 CE Kennzeichnung Symbol for the marking of electrical and electronics devices according to Directive 2002 96 EC Direct current Non ionizing radiation USB Plug ...

Страница 5: ...or to convert internal bodysounds such as heart and lung sounds to electronic Signals and deliver to suitable PC or mobile device 1 4 1Indication The instruments serve the trained doctor or specialist as an aid in the detec tion diagnosis monitoring and treatment the alleviation of illnesses injuries or disabilities 1 4 2 Contraindication There are no contraindications 1 4 3 Intended patient popul...

Страница 6: ...d part This product is not suitable for use in the presence of a flammable anesthetic mixture with air or with continuous oxygen or nitrous oxide Allergic reaction to the skin Only use the stethoscope when it has been cleaned Do not use store or transport this product in excessively dusty conditions There are no modifications to this product that a user can make and they should not attempt to do s...

Страница 7: ...d person onment nesthetic s y should r Repre c PCP 1 Art Nr 4310 ri sonic PCP 1 Instruction of use 2 2 Device Function ri sonic PCP USB connector with straight cable ri sonic PCP 1 with spiral cable Chest Piece Top ...

Страница 8: ... is intended for use with a standard PC mobi le devices with USB port and appropriate app or software which is not included The ri sonic provides sound data as specified The user must check whether the line data of the sensor matches the requirements of the PC mobile device app or software The ri sonic PCP USB Stethoscope consists of a hardware element the PCP Chest Piece The PCP USB Chest Piece c...

Страница 9: ...e chest piece assembly along with the piezo element PE The following figure shows a cross section of a physical representation of ri sonic PCP USB Chest Piece assembly ri sonic PCP 1 The ri sonic PCP 1 stethoscope is intended for use with a standard PC mobile devices with Microphone port and appropriate app or software which is not in cluded The ri sonic brings its performance data The user must c...

Страница 10: ...y coverage 20 2000 Hz 3 2 Commissioning 3 2 1 Installation ri sonic PCP USB Installation of the ri sonic PCP USB Stethoscope is comprised of plugging the USB connector of the ri sonic PCP USB Chest Piece into the USB port of the PC For usage with PC only no additional software ist required For use with speciali zed software follow the instructions of the Software 3 2 2 Installation ri sonic PCP 1 ...

Страница 11: ...efore sending back defective products for repair the following instructions should be followed 4 2 Cleaning and Disinfection In order to avoid possible cross contamination the stethoscope must be cleaned and disinfected regularly before each use The stethoscopes can be cleaned on the outside using a damp cloth if necessa ry moistened with alcohol until optical cleanliness is achieved Use disinfect...

Страница 12: ...ions 25 C bis 35 C 35 C to 70 C at a water vapor pressure up to 50mbar a relative humidity range of 0 to 90 Airpressure 700hPa to 1060hPa There is no minimum time required for the unit to warm up from minimum sto rage temperature or to cool down from maximum storage temperature before use If a unit is exposed to environmental conditions outside these ranges it should be checked for proper operatio...

Страница 13: ...c Compatibility EMC electromagnetic compatibility During installation and operation of the device observe the following instruc tions Do not use the device simultaneously with other electronic equipment to avoid electromagnetic interference with the operation of the device Do not use or stack the device near on or under other electronic equipment to avoid electromagnetic interference with the oper...

Страница 14: ...earby devices It may become necessary to take appropriate corrective measures such as redirecting or rearranging the ME device or shield The rated ME device does not exhibit any basic performance features in the sense of IEC 60601 1 which would present an unacceptable risk to patients operators or third parties should the power supply fail or malfunction Warning Portable stationare RF communicatio...

Страница 15: ...nvironment Emissions test Compliance Electromagnetic environment guidance RF emissions CISPR 11 Group 1 The Electronic Stethoscopes ri sonic PCP USB ri sonic PCP 1 uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF emissions CISPR 11 Class B The Electronic Stethoscopes ri sonic PCP USB...

Страница 16: ...ble and mobile RF commu nications equipment should not be used closer to any part of the Electronic Stethoscopes ri so nic PCP USB ri sonic PCP 1 including the cables than the recommended distance which is calculated using the equation applicable to the transmitter fre quency Recommended separation dis tance d 1 2 P 150 KHz to 80 MHz d 1 2 P 80 MHz to 800 MHz d 2 3 P 800 MHz to 2 7 GHz Where P is ...

Страница 17: ...ri sonic PCP 1 is used exceeds the applicable RF compliance le vel above the Electronic Stethoscopes ri sonic PCP USB ri sonic PCP 1 should be observed to verify normal operation If abnormal performance is observed additional measures may be necessary such as re orienting or relocating the Electronic Stethoscopes ri sonic PCP USB ri sonic PCP 1 b Over the frequency range 150 kHz to 80 MHz field st...

Страница 18: ...maximum output power of the communications equipment Rated maxi mum output power of the trans mitter W Separation distance according to the frequency of the transmitte m 150 kHz 80 MHz 80 MHz 800 MHz 800 MHz 2 7 GHz 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maximum output power not listed above the re commended separation distance ...

Страница 19: ...s does not apply to wearing parts A war ranty claim can only be granted if this warranty card has been completed and stamped by the dealer and is enclosed with the product Please remember that all warranty claims must be made during the warranty period We will of course be pleased to carry out checks or repairs after expiry of the warranty period at a charge You are also welcome to request a provi...

Страница 20: ...20 Rudolf Riester GmbH P O Box 35 Bruckstraße 31 DE 72417 Jungingen Germany Tel 49 7477 9270 0 Fax 49 7477 9270 70 E Mail info riester de www riester de 99449 Rev 0 2021 06 Subject to alterations ...