Guidelines and manufacturer‘s declaration – electromagnetic immunity
The ri-focus LED headlights and clar N LED head mirrors are intended for use in the electromagnetic environment specified below. The customer or user of the
headlight should ensure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance
Electromagnetic environment - guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
Con: ±8 kV
Air: ± 2, 4, 8,
15 kV
Con: ±8 kV
Air: ± 2, 4, 8,
15 kV
Floors should be wood, concrete or ceramic tile. If the floor is covered
with synthetic material, the relative humidity must be at least 30%.
Rapid transients electri-
cal disturbances/bursts
IEC 61000-4-4
Not applicable
Not applicable
The quality of the supply voltage should be that of a typical commercial
or hospital environment.
Impulse voltage
IEC 61000-4-5
Not applicable
Not applicable
The quality of the supply voltage should be that of a typical commercial
or hospital environment.
Voltage dips, short
interruptions and voltage
variations according to
IEC 61000-4-11
Not applicable
Not applicable
The quality of the supply voltage should be that of a typical com-
mercial or hospital environment.
Magnetic field with
energy-efficient rated
frequencies
IEC 61000-4-8
30A/m
30A/m
Magnet fields at mains frequency should be at a level typical for the
location of a typical commercial hospital environment.
Note: UT is the AC source. Mains voltage before the application of the test level.
Содержание ri-focus LED
Страница 73: ...73 B MD II 2 REF C F 2002 96 EC 1 2...
Страница 74: ...74 1 3 IEC 60601 1 2 1 4 Riester 1 4 1 6 6 1 4 2 Riester 1 4 3...
Страница 75: ...75 1 4 4 1 4 5 1 4 6 1 5 Riester...
Страница 77: ...77 2 3 2 3 1 2 3 2 ri focus 1 1 90 clar N 55 2 3 3 ri focus clar N AAA CR 123A 1 1 1...
Страница 82: ...82 8 1 IEC 60601 1 30 12...
Страница 83: ...83 ri focus Clar N CISPR 11 1 CISPR 11 B IEC 61000 3 2 IEC 61000 3 3...
Страница 87: ...87 8 1 9 2 B RIESTER Rudolf Riester GmbH Repairs dept RR Bruckstr 31 D 72471 Jungingen Germany...