6.3.6 Several Additional Commands
6.3.6.1 Getting Version, Type and Serial Number Information
Command
Meaning
. #V<Cr>
Getting Software Version
. #VI<Cr>
Getting Instrument Ident
. #VN<Cr>
Getting Instrument Name
. #VM<Cr>
Getting Instrument Modification
. #SN<Cr>
Getting Instruments serial number
6.3.6.2 Getting Supply Voltage and Instrument Temperature
Command
Meaning
. V<Cr>
Getting supply voltage in units of 0.1 volt. This voltage is measured internally,
therefore voltage loss from the supply cable can not be taken into account.
. TEMP<Cr>
Getting Temperature within instrument in units of degree Celsius. Note that
the temperature within in the instrument usually is some degrees higher than
ambient temperature.
6.3.6.3 Error Handling
Command
Meaning
. ERR<Cr>
Get a list of errors. Each error detected is listed in a line, where each line
starts with “\” and is followed by an error message. The last line then is
“=ERR” as response to the “.ERR” command.
Note that while in measurement mode errors are automatically sent as
messages.
ERRACK<Cr>
Error acknowledge. This command clears all errors listed with .ERR and no
errors are pending any more. Note that FATAL errors can not be
acknowledged.
For detailed description of error handling and error lists see chapter 6.6