ADC Calibration
PC calibration
• After replacing the optical engine / main board assembly, do the following steps.
• If you do not have professional equipment such as Quantum Data 802B or CHROMA2327, use a
PC that supports HDTV resolution & has an independent graphics card to output the corresponding
PC pattern. You can download the “test pattern by PC” from the website as shown below.
Procedure
• Test equipment: Video generator.
• Test signal: Analog 1280 x 800@60Hz
• Test Pattern: 94% White(up)/6% Black (down) [A]
• Calibration pattern should be in full screen.
1.
Get into Service Mode.(
page 53 "How to Enter the Service Mode"
)
5. Test & Inspection
60
Содержание Polaris-PJ1
Страница 2: ......
Страница 7: ...1 Product Information Overview Main Unit 1 Zoom lever 2 Control panel 3 Connection ports 5...
Страница 10: ...The interface is subject to model s specifications 1 Product Information 8...
Страница 22: ...2 Installation 20...
Страница 71: ...4 Click Next 5 Click Next 6 Click Next Firmware Update 69...
Страница 81: ...MEMO 79...
Страница 82: ...MEMO 80 EN...