EDP Commands

© Rice Lake Weighing Systems

 

 All Rights Reserved

31

5.1.6 Normal Mode Commands

The serial transmit weight data commands (see 

Table 5-2 on page 28

) transmit data to the EDP port on demand. The SX and 

EX commands are valid only in normal operating mode; all other commands are valid in either setup or normal mode.

5.2 Saving and Transferring Data

Connecting a personal computer to the Tracer AV EDP port allows you to save configuration data to the PC or to download 

configuration data from the PC to a Tracer AV. The following sections describe the procedures for these save and transfer 

operations.

5.2.1 Saving Tracer AV Data to a Personal Computer

Configuration data can be saved to a personal computer connected to the EDP port. The PC must be running a 

communications program such as PROCOMMPLUS

®

. See 

Section 2.3.3 on page 10

 for information about serial 

communications wiring and EDP port pin assignments.
When configuring the Tracer AV, ensure that the values set for the BAUD and BITS parameters on the SERIAL menu match the 

baud rate, bits, and parity settings configured for the serial port on the PC. Set the PRNDEST parameter to EDP.
To save all configuration data, place the Tracer AV in setup mode and send the DUMPALL EDP command. The Tracer AV 

responds by sending all configuration parameters to the PC as ASCII-formatted text.

5.2.2 Downloading Configuration Data from PC to Tracer AV

Configuration data saved on a PC or floppy disk can be downloaded from the PC to a Tracer

 

AV. This procedure is useful when 

a number of Tracer

 

AVs with similar configurations are set up or when a Tracer AV is replaced.

To download configuration data, connect the PC to the EDP port as described in 

Section 5.2.1

. Place the Tracer AV in setup 

mode and use the PC communications software to send the saved configuration data to the 

Tracer AV

. When transfer is 

complete, calibrate the Tracer

 

AV as described in 

Section 4.0 on page 24

.

 

• Calibration settings are included in the configuration data downloaded to the Tracer AV. If the receiving Tracer AV is a direct replacement 

for another Tracer AV and the attached scale is not changed, recalibration is not required.

• When downloading configurations that include changed serial communications settings, edit the data file to place the serial 

communications changes at the end of the file. Communication between the PC and Tracer AV will be lost once the Tracer AV receives 

settings for baud rate (BAUD parameter) or data bits and parity (BITS parameter) that do not match those configured for the PC.

Command

Description

Response Format

UID

Set unit ID

nnnnnnn

SD

Set date

MMDDYY, DDMMYY, YYMMDD (enter using DATEFMT specified)

ST

Set time

hhmm (enter using 24-hour format)

SX

Start EDP streaming

OK 

or

 ??

EX

Stop EDP streaming

OK 

or

 ??

RS

Reset system

XG

Transmit gross weight in displayed units

nnnnnn UU

where 

nnnnnn

 is the weight value, 

UU

 is the units.

XN

Transmit net weight in displayed units

XT

Transmit tare weight in displayed units

XG2

Transmit gross weight in non-displayed units

XN2

Transmit net weight in non-displayed units

XT2

Transmit tare weight in non-displayed units

XE

Query system error conditions

nnnnn nnnnn

See 

Section 6.1.2 on page 32

 for detailed information about the XE com

-

mand response format.

Table 5-11.  Normal Mode Commands

Note

Содержание Tracer AV

Страница 1: ...Tracer AV Aviation Baggage Instrumentation Installation Manual PN 98316 Rev B October 25 2019...

Страница 2: ...marks of their respective companies All information contained within this publication is to the best of our knowledge complete and accurate at the time of publication Rice Lake Weighing Systems reserv...

Страница 3: ...ications 10 2 3 4 Digital I O 10 2 4 Board Removal 11 2 5 Replacement Parts 11 3 0 Configuration 12 3 1 Configuration Methods 12 3 1 1 EDP Command Configuration 12 3 1 2 Front Panel Configuration 12 3...

Страница 4: ...5 2 2 Downloading Configuration Data from PC to Tracer AV 31 6 0 Appendix 32 6 1 Error Messages 32 6 1 1 Displayed Error Messages 32 6 1 2 Using the XE EDP Command 32 6 2 Status Messages 33 6 2 1 Usin...

Страница 5: ...the warnings could result in injury or death Contact any Rice Lake Weighing Systems dealer for replacement manuals Failure to heed could result in serious injury or death Ensure the power cord is disc...

Страница 6: ...remove the setup switch access screw on the Tracer AV See Figure 2 3 on page 5 Once the screw is removed use a small screwdriver or pen tip to press the setup switch The display changes to show the w...

Страница 7: ...at standstill or within the specified motion band Some operations including tare function can only be done when the standstill symbol is shown lb and kg annunciators indicate the units associated wit...

Страница 8: ...V to ensure all components are included and undamaged The shipping carton should contain the Tracer AV enclosure remote display remote keypad this manual two cables and a power cord If any parts were...

Страница 9: ...for the power cord one to accommodate the load cell communications remote displays and keypads Figure 2 3 shows the assignments for the Tracer AV connections Details on pin assignments are shown in T...

Страница 10: ...TVS1 TVS5 TVS4 3 3 R7 R4 C16 C4 DIGIN 1 J1 LOAD CELL U2 D1 U19 U6 U9 SERIAL 1 9 1 1 8 1 1 Gn d 20mA Gnd TxD Gnd Gn d DI 2 EXC E XC SEN SEN SIG 5 1 1 1 20mA RxD RxD TxD DI 1 SIG Micro processo r DISPLA...

Страница 11: ...s Reserved 7 Figure 2 6 Panel Mount Housing Enclosure Dashed line represents RD 2 display location LCD Solid line represents RD 1 display location LED Figure 2 7 Remote Display Cutout for Panel Mount...

Страница 12: ...Enclosure Dashed line represents RD 2 display location LCD for External Mounted Enclosure Solid line represents RD 1 display location Figure 2 9 Remote Display Cutout for External Mount 7 50 3 90 LCD...

Страница 13: ...oad cell This DB 9 is connected to the CPU board on J1 see the table below for the pin assignments The DB 9 is for four wire installations so leave sense jumper JP1 and JP2 on The DB 9 is factory pinn...

Страница 14: ...Digital outputs are typically used to control relays that drive other equipment In the Tracer AV the setpoint value will cause the display to flash Outputs are designed to sink not source switching c...

Страница 15: ...on page 6 for connector locations 6 Remove the five screws from the CPU board then lift the board out of the enclosure To replace the CPU board reverse the above procedure 2 5 Replacement Parts Table...

Страница 16: ...ront panel and provide some functions not otherwise available EDP commands can be used to simulate pressing front panel keys to configure the Tracer AV or to dump lists of parameter settings See Secti...

Страница 17: ...the six digit display 2 Press Zero to accept the modified value 3 1 5 Editing Floating Point Decimal Point Position 1 By pressing Units navigate all the way to the right beyond all characters until th...

Страница 18: ...Menu Figure 3 3 Default Calibration Menu DEFCAL Menu Parameter Choices Description SELECT 500 lb 300 lb 500 lb sets the scale to a primary of 500 lb x 1 lb and a secondary of 250 kg x 5 kg SP1 trip 50...

Страница 19: ...y divisions Maximum legal value varies depending on local regulations Note For scales using linear calibration do not set the zero tracking band to a value greater than that spec ified for the first l...

Страница 20: ...rom the immediate area of the scale Choices indicate the number of A D conversions that are averaged to obtain the displayed reading A higher number gives a more accurate display by minimizing the eff...

Страница 21: ...nits See Level 3 submenu parameter descriptions DSPRAT 250MS 500MS 750MS 1SEC 1 5SEC 2SEC 2 5SEC 3SEC 4SEC 6SEC 8SEC Display rate Sets the update rate for displayed values Values are in milliseconds M...

Страница 22: ...8 888 888 8888 88 Decimal point location Determines the location of the decimal point or dummy zeros in the secondary unit dis play DSPDIV 5D 1D 2D Display divisions Selects the value of minimum divis...

Страница 23: ...has been set WVAL Display and edit the test weight value WSPAN Display and edit the span calibration A D count value REZERO Press Zero to remove an offset value from the zero and span calibrations Us...

Страница 24: ...ERMIN CR LF CR Termination character Selects termination character for data sent from the EDP port EOLDLY 000 number End of line delay Sets the delay period in 0 1 second intervals from when a formatt...

Страница 25: ...ss or net mode as long as the current weight is within the specified ZRANGE In OIML mode the scale must be in gross mode before it can be zeroed pressing the ZERO key in net mode clears the tare UID 1...

Страница 26: ...and value used by the setpoint Level 3 submenus ENABLE OFF ON Turns setpoint ON or OFF KIND GROSS NET Specifies the setpoint kind and determines whether function is based on GROSS or NET weight GROSS...

Страница 27: ...f the band value BNDVAL number The band value for either INBAND or OUTBND trip setpoints BNDVAL is ignored unless trip is set to INBAND or OUTBND HYSTER number Specifies a band around the setpoint val...

Страница 28: ...n Figure 4 2 Default Calibration DEFCAL Menu 4 1 Default Calibration If using a 500 lb or 300 lb scale you can select it from the DEFCAL menu This automatically uses the settings shown in Table 4 1 fo...

Страница 29: ...s to show the word DEFCAL 2 Remove all weight from the scale platform If test weights require hooks or other items place them on the scale for zero calibration 3 Press Units until the display reads CA...

Страница 30: ...re About Calibration The following topics provide additional information about compensating for environmental factors Section 4 4 1 and diagnostic information for determining expected zero and span co...

Страница 31: ...ax used for each of these groups 5 1 1 Key Press Commands Key press EDP commands Table 5 1 simulate pressing the keys on the 2 button keypad Most commands can be used in both setup and weighing mode S...

Страница 32: ...yed in either setup mode or normal mode using the following syntax COMMAND ENTER Most parameter values can be changed in setup mode only Use the following command syntax when changing parameter values...

Страница 33: ...8 888888 888880 PRI DSPDIV Primary units display divisions 1D 2D 5D PRI UNITS Primary units LB KG OZ TN T G NONE SEC DECPNT Secondary units decimal position 8 88888 88 8888 888 888 8888 88 88888 8 888...

Страница 34: ...ALUE SP2 VALUE Setpoint value number SP1 TRIP SP2 TRIP Trip Higher Lower Inband Outbnd SP1 BNDVAL SP2 BNDVAL Band value number SP1 HYSTER SP2 HYSTER Hysteresis number SP1 ACCESS SP2 ACCESS Setpoint ac...

Страница 35: ...r AVs with similar configurations are set up or when a Tracer AV is replaced To download configuration data connect the PC to the EDP port as described in Section 5 2 1 Place the Tracer AV in setup mo...

Страница 36: ...ge Description Solution E A D A D physical error Contact Rice Lake Weighing Systems EEEROM EEPROM physical error EVIREE Virgin EEPROM Use TEST menu to perform DEFLT restore defaults procedure then rec...

Страница 37: ...6 3 If more than one annunciator is lit the second number returned is the sum of the values representing the active annunciators Example If the annunciator status value returned on the ZZ command is 1...

Страница 38: ...S Text or 100DMText Figure 6 1 Continuous Output Data Format Note STX POL wwwwwww UNIT G N S TERM ASCII 02 decimal Weight data 7 digits right justi ed with decimal point leading zero supression Overlo...

Страница 39: ...Front Panel Display Characters Figure 6 2 shows the 7 segment LED character set used to display alphanumeric characters on the Tracer AV front panel Figure 6 2 Tracer AV Display Characters 9 E Q F R...

Страница 40: ...cimal point is set to 8888 88 the indicator will overflow if 5 tons or more are applied to the scale With 5 tons applied and a conversion factor of 2000 the secondary unit s display needs five digits...

Страница 41: ...iltering effect is 20 4 8 8 With this configuration each A D reading has a 1 in 20 effect on the displayed weight value Setting the filters to 1 effectively disables digital filtering 6 6 2 DFSENS and...

Страница 42: ...le in Step 2 with a threshold weight value of 50 lb and a display division value of 5 lb 50 5 lb 10 DD DFTHRH should be set to 10 DD for this example 5 Finally set the DFSENS parameter high enough to...

Страница 43: ...ero or negative no TARE ZERO yes CLEAR TARE positive no TARE yes CLEAR TARE Table 6 6 TARE and ZERO Key Functions for REGULAT Parameter Settings Serial Port Status LED1 EDP Port Flashing red TXD activ...

Страница 44: ...inimum at 50 or 60 Hz Input Overload 12 V continuous static discharge protected RFI Protection Signal excitation and sense lines protected by capacitor bypass Analog Output Optional fully isolated vol...

Страница 45: ......

Страница 46: ...nada Mexico 800 321 6703 International 715 234 9171 Europe 31 0 26 472 1319 Rice Lake Weighing Systems is an ISO 9001 registered company Rice Lake Weighing Systems Specifications subject to change wit...

Отзывы: