RoughDeck Low Profile Floor Scale

2.4.2

Mounting Plate Installation

For permanent applications, the scale should be secured to the floor to prevent sideways movement. Two mounting 
plates, with holes that slightly exceed the foot diameter, are available as an option for that purpose.

Lift the scale so that the feet are approximately one inch off the floor. Slide mounting plates under two diagonally 
opposed feet. Lower the scale back to the floor, and position the plates as shown in Figure 2-2 so that the boltdown 
holes are accessible from above. 

Using the mounting plates as templates, drill pilot holes into the floor for suitable anchor bolts. Bolt the plates to 

the floor using 

1

/

2

-in anchor bolts (

3

/

4

-in bolts for 20K and 30K-lb capacity scales). Recheck foot adjustment and 

deck level after this operation. 

For installations using access ramps, mounting plates are not necessary as the ramps have built-in mounting plates 
to secure the scale feet

3.50" (2K – 20K models); 4.50" (30K)

.562" DIA (2K–10K models)

.812" DIA (20K & 30K 

MOUNTING PLATE

SCALE DECK

FOOT

Figure 2-2. Mounting Plate Installation

2.5

Electrical Interface to the Indicator

Twenty feet of 6-wire cable to connect the scale to the weight indicator is supplied with each scale.

NEMA 4X

If your scale uses the NEMA 4X junction box, push one end of this cable into the junction box through the strain 
relief bushing hole in the tray. Connect the wires to the INDICATOR terminal (shaded area of Figure 2-3) as shown 
in the load cell specifications. Pull out excess and tighten the strain relief bushing to hold the cable snugly.

Standard Junction Box

If using the standard junction box, put the cable through the hole in the junction box tray. Slide the two parts of the 
strain relief flex fitting onto the cable with the threaded piece inside the faceplate and the flexible spiral portion to 
the outside. Push the cable end into the junction box and connect the wires to the INDICATOR terminal. Pull out 
excess cable and tighten the two parts of the strain relief flex fitting to hold the cable snugly.

Содержание RoughDeck ALTRALITE

Страница 1: ...RoughDeck Low Profile Floor Scale SS HE HP HP H HC and ALTRALITE Installation Manual utilizes economical junction box 16960 Rev B...

Страница 2: ......

Страница 3: ...4 2 Periodic Maintenance 13 4 3 Load Cell Replacement 13 4 3 1 Load Cell Wiring to the Junction Box 15 4 4 Replacement Parts List and Accessories 16 4 4 1 Pit Frames 16 4 4 2 Access Ramps 16 4 4 3 Rep...

Страница 4: ...ii RoughDeck Rice Lake continually offers web based video training on a growing selection of product related topics at no cost Visit www ricelake com webinars...

Страница 5: ...losed in conduit through the main channels and held down with replaceable cable ties near each corner eliminating the possibility of cable damage in portable applications Also useful for portable appl...

Страница 6: ...r death Contact any Rice Lake Weighing Systems dealer for replacement manuals Proper care is your responsibility Before attempting to operate this unit make sure every individual who operates or works...

Страница 7: ...xternal mounted NTEP certified model with hermetically sealed load cells and NEMA 4X J box for hostile environments 1 3 Operating Requirements Electrical Grounding For systems where the scale is conne...

Страница 8: ...e where overweight loads can maneuver easily without crossing the platform Avoid areas where damage could occur from side impacts of wheels or forklift tines Avoid areas where falling objects could ca...

Страница 9: ...curely attached through the nuts welded to the bottom side of the top plate Lifting from the bottom of the plate could cause nuts to break loose and the scale to fall Figure 2 1 Proper Lifting Techniq...

Страница 10: ...s to secure the scale feet 3 50 2K 20K models 4 50 30K 562 DIA 2K 10K models 812 DIA 20K 30K MOUNTING PLATE SCALE DECK FOOT Figure 2 2 Mounting Plate Installation 2 5 Electrical Interface to the Indic...

Страница 11: ...scale for servicing or cleaning LOAD CELL CABLE STEEL COVER FLOOR SCREW LOAD CELL CABLE STEEL COVER Figure 2 4 Cable Protection Strain Relief Flex Fitting Figure 2 5 Standard Junction Box Note In wash...

Страница 12: ...loor sweepings or material spills may accumulate in the pit and interfere with scale operation periodically clean the pit All RoughDeck models have threaded holes for eye bolts so the scale can be eas...

Страница 13: ...sary Access ramps can only be attached to the scale on one of the two scale sides that are perpendicular to the longitudinal axis of the load cells For example the scale shown in Figure 2 8 could have...

Страница 14: ...ll must be matched by adjusting the signals with potentiometers at the junction box a process known as trimming All assembled RoughDeck scales are delivered with the junction box corner trimmed Corner...

Страница 15: ...of the test weight used 3 3 Calibration Procedure Refer to the indicator manual to determine correct calibration procedures It is recommended that the scale be exercised before calibration to be certa...

Страница 16: ...brated Calibrate according to indicator manual and Section 3 3 on page 11 Load cells faulty Test and replace load cells if necessary Feet touching deck underside Adjust feet downward to provide cleara...

Страница 17: ...ELL 2 CELL 1 1 4 2 3 Note Figure 4 2 shows both the scale and the junction box as viewed from the bottom To verify correct load cell junction box terminal matching see the numbers on the terminals ins...

Страница 18: ...n trimming and calibration is necessary after load cell replacement Follow instructions in Section 3 2 and Section 3 3 on page 11 4 3 1 Load Cell Wiring to the Junction Box 1 4 2 3 Figure 4 4 Corner N...

Страница 19: ...764 39599 39612 5 x 6 1 53 x 1 83 18769 18770 39600 5 x 7 1 53 x 2 13 18771 18772 39601 39614 6 x 6 1 83 x 1 83 39602 39615 6 x 8 1 83 x 2 44 39604 7 x 9 2 13 x 2 74 39605 7 x 10 2 13 x 3 05 39606 396...

Страница 20: ...ength x Ramp Height Access Ramp for Model ft x ft x inch m x m x mm HP H 20K HP H 30K 4 x 4 x 4 1 22 x 1 22 x 102 39585 5 x 4 x 4 1 52 x 1 22 x 102 39586 6 x 4 x 4 1 83 x 1 22 x 102 39587 7 x 4 x 4 2...

Страница 21: ...bber grommet load cell cable HP HP H HC models Rubber grommet deck to indicator cable HP HP H H HC models 8 15658 Cable tie mount 1 square 9 15631 Cable ties 3 10 15071 15075 40337 Load cell screw HP...

Страница 22: ...that the nonconformity actually exists and was not caused by accident misuse neglect alteration improper installation improper repair or improper testing RLWS shall be the sole judge of all alleged no...

Страница 23: ...9171 Immediate Emergency Service For immediate assistance call toll free 1 800 472 6703 Canadian and Mexican customers please call 1 800 321 6703 If you are calling after standard business hours and h...

Страница 24: ...20 RoughDeck Low Profile Floor Scale...

Страница 25: ......

Страница 26: ...8 USA U S 800 472 6703 Canada Mexico 800 321 6703 International 715 234 9171 Europe 31 0 26 472 1319 www ricelake com www ricelake mx www ricelake eu www ricelake co in m ricelake com Rice Lake Weighi...

Отзывы: