![Rice Lake RL-TA Series Скачать руководство пользователя страница 36](http://html1.mh-extra.com/html/rice-lake/rl-ta-series/rl-ta-series_operation-manual_1454207036.webp)
32 RL-TA Series Operation Manual
Obtain a total weight without reloading the balance
Press the Zero/Tare key in step 3 without unloading the balance. This will reset the
display to “0”.
Put additional samples on, and press the Target key. Then, the weight is cumulated
.
1.
Put additional samples on when the display indicates “0”.
2.
When [
t
-
Err
] is displayed when you press the Target key, it indicates that you put additional
samples on twice, that you unloaded some samples, or that you pressed the key without adding
samples.
3.
You can add samples when an asterisk (*) is displayed.
4.
The [
J. tA.
] display with Function 1 can be used for turning ON/OFF the function to wait for
stabilization upon additions.
12.2
Net Addition Function
The net addition function automatically performs tare range after an addition, so additions can be
performed without having to replace samples.
Function 1 is [
2. SEL 1
] and [
26 Ad. N 2
].
1.
Reset the display to “0”.
Press the Zero/Tare key.
2.
Put samples on the balance, and obtain the weight.
Put samples to weigh on the balance. When the asterisk (*)
sign is displayed, you can put additional samples on.
When
is displayed, press the
Target key
. The
displayed weight is added, and the cumulative
weight is displayed with a [
Σ
] sign for several seconds.
3.
The balance automatically sets tare range.
The balance automatically sets the tare range.
4.
Put additional samples on the balance to obtain the weight.
When
is displayed after adding the sample, press
the
Target key
.
The displayed value is added, and the cumulative weight
is displayed with a [
Σ
] sign for several seconds.