background image

Setup

© Rice Lake Weighing Systems 

 All Rights Reserved

5

3.2

Initial Setup

Use the following steps to set up and verify the RL-DBS-2 upon initial power up and periodically thereafter.

1. Plug the AC adapter cable into the back of the scale (

Figure

 

2-1 on page

 

3

) and the AC adapter plug into a power 

source.

2. Turn on the scale. A self test runs and 0.00.0 lb or 0.00 kg displays when complete. 
3. Place a weight, not exceeding 40 lb, on the scale. The weight measurement is complete when 

o

 displays. 

4. Press 

LB/KG

 to select unit of measure, lb/oz or kg.

5. Remove the weight from the scale. To display the weight again, press 

RECALL

Note

.

  If setup procedure fails, see 

Section

 

6.3 on page

 

9

. If the problem is not resolved, refer to a qualified service provider.

3.3

Set Units

The RL-DBS-2 can be set up to display weights in lb only, kg only or to toggle between lb and kg units on the scale.

3.3.1

KG Units Only

Lock the weight units to display in kg only.

1. Ensure the scale is turned on.

2. Press 

LB/KG

 to change the displayed units to kg.

3. Press and hold 

LB/KG

 for five seconds until kg flashes.

4. Release 

LB/KG

. The weight units are now locked in kg. Pressing 

LB/KG

 does not switch units to lb.

3.3.2

LB Units Only

Lock the weight units to display in lb only.

1. Ensure the scale is turned on. 

2. Press 

LB/KG

 to change the displayed units to lb.

3. Press and hold 

LB/KG

 for five seconds until lb oz flashes.

4. Release 

LB/KG

. The weight units are locked in lb. Pressing 

LB/KG

 does not switch units to kg.

3.3.3

Unlock Units

Unlock the weight units to toggle between lb and kg. 

1. Ensure the scale is turned on.

2. Press and hold 

LB/KG

 for five seconds until kg or lb oz flashes.

3. Release 

LB/KG

. Pressing 

LB/KG

 toggles units between lb and kg.

Содержание RL-DBS-2

Страница 1: ...RL DBS 2 Digital Baby Scale Technical Manual PN 168522 Rev D July 5 2018...

Страница 2: ...marks of their respective companies All information contained within this publication is to the best of our knowledge complete and accurate at the time of publication Rice Lake Weighing Systems reserv...

Страница 3: ...4 9171 and asking for the training department 1 0 Introduction 1 1 1 Safety 2 2 0 Assembly 3 3 0 Setup 4 3 1 Display Descriptions 4 3 2 Initial Setup 5 3 3 Set Units 5 4 0 Operation 6 4 1 With a Tare...

Страница 4: ...2 Digital Baby Scale ii Visit our website www RiceLake com Rice Lake continually offers web based video training on a growing selection of product related topics at no cost Visit www ricelake com web...

Страница 5: ...y spontaneous movement of the baby The weight can be displayed in pounds ounces or in kilograms This manual can be viewed and downloaded from the Rice Lake Weighing Systems website at www ricelake com...

Страница 6: ...ighing Systems dealer for replacement manuals Failure to heed could result in serious injury or death Ensure every individual operating or working with this unit has read and understands the following...

Страница 7: ...ale on a sturdy level surface 2 Connect the 120 V or 230 V AC adapter to the back of the scale AC Power Connection 120V Adapter 230V Adapter Figure 2 1 AC Power Connection Site The RL DBS 2 can also b...

Страница 8: ...stable and ready for use 4 Memory annunciator Indicates the weighing process is complete 5 6 Digit LCD digital display Displays the weight in pounds ounces or kilograms 6 On Off Press to turn On Press...

Страница 9: ...its The RL DBS 2 can be set up to display weights in lb only kg only or to toggle between lb and kg units on the scale 3 3 1 KG Units Only Lock the weight units to display in kg only 1 Ensure the scal...

Страница 10: ...d on the weighing tray ensure the edges of the pad are not touching the stand o displays when the weight reading is complete 3 Press ON TARE to remove the weight of the pad or any other accessory on t...

Страница 11: ...y 6 Press ON TARE again C20 briefly displays and lb is lit then 0 00 displays If calibrating in kg C10 displays and kg is lit then 0 00 displays 7 Load the scale with a 20 lb 10 kg test weight 8 Press...

Страница 12: ...ion 4 Press and hold LB KG and ON TARE at the same time until CAL displays 5 Release the keys The raw A D counts display 6 Press ON TARE 0 00 displays 7 Load the scale with a 40 lb or 20 kg test weigh...

Страница 13: ...cloth 6 3 Troubleshooting Use the following instructions to troubleshoot any errors with the scale Table 6 1 Troubleshooting Table Symptom Possible Cause Corrective Action Scale does not turn on Batt...

Страница 14: ...22 44 lb x 0 5 oz 10 20 kg x 10 g Power Requirements Four AA batteries included AC adapter included Adapter is UL and C UL approved Environmental Operating temperature 50 to 104 F 10 to 40 C Storage...

Страница 15: ...RLWS confirms that the nonconformity actually exists and was not caused by accident misuse neglect alteration improper installation improper repair or improper testing RLWS shall be the sole judge of...

Страница 16: ...RL DBS 2 Digital Baby Scale 12 Visit our website www RiceLake com...

Страница 17: ......

Страница 18: ...anada Mexico 800 321 6703 International 715 234 9171 Europe 31 0 26 472 1319 Rice Lake Weighing Systems is an ISO 9001 registered company Rice Lake Weighing Systems Specifications subject to change wi...

Отзывы: