Troubleshooting and Maintenance

© Rice Lake Weighing Systems

 

 All Rights Reserved

11

4.0

Troubleshooting and Maintenance

4.1 Troubleshooting

Problem

Possible Cause

Solution

Display is blank when POWER 

key is pressed

Discharged battery

Replace cells, or if using NiMH, recharge

Defective battery

Replace

Corroded battery or battery contacts

Clean contacts

Defective switch or circuit board

Requires authorized service

Display does not function 

properly or front panel keys do 

not function normally or 

Dyna-Clamp not powering off

Improperly loaded software

Reinstall software

Faulty circuit board

Requires authorized service

Loose connectors

Requires authorized service

Dyna-Clamp does not respond 

to tension changes

Out of calibration

Requires authorized service

Faulty load cell

Requires authorized service

Worn out or damaged drive rod

Requires authorized service

Display over ranges below 100% 

of capacity

Zero requires adjustment

Rezero the Dyna-Clamp

Too much tension/load has been zeroed

Rezero the Dyna-Clamp

Display drifts

AZM (Auto0) is turned off

Turn AZM on

Rapid temperature changes such as moving 

the Dyna-Clamp from indoors to outdoors

Wait until the Dyna-Clamp temperature has stabilized

Displayed tension shows larger 

error

Dyna-Clamp not zeroed before load is lifted Zero the Dyna-Clamp with no load attached
lbf/kgf/kN units causing confusion

Select proper units

Requires recalibration

Requires authorized service

Display reading not stable

Excessive vibration in wire rope

Increase filtering

Load cell faulty

Check LC connections

Display toggles between Error 

and Load

Tension exceeds capacity

Immediately reduce tension

Faulty load cell or wiring

Check LC and LC wiring

Display toggles between Error 

and Key

A key is stuck or is being held down

Check switches for damage
Ensure that a remote is not continuously transmitting

Low battery is blinking

Battery is low

Replace (alkaline) or recharge batteries

Unit turns on then immediately off Battery is low

Replace (alkaline) or recharge batteries

Tension does not zero

System not stable

Wait for standstill annunciator to turn on
Increase filtering for more stability

Zero out of range

Zero range might be limited; only zero while Dyna-Clamp is not 

clamped to a wire rope

Tension does not average

System is not stable

Wait for standstill annunciator to turn on, or see Display reading not 

stable above for an additional solution

Low threshold % of capacity not met between 

average measurements

Remove Dyna-Clamp from the rope being tested, the device must drop 

below the low % of capacity before a new average can be taken

Setpoint LEDs blink

Range is enabled and a trigger point has 

been reached

Disable range if it is not needed

Table 4-1.  Troubleshooting

Содержание MSI Dyna-Clamp

Страница 1: ...MSI Dyna Clamp Tension Meter Operation Manual May 19 2020 PN 193853 Rev A...

Страница 2: ...marks of their respective companies All information contained within this publication is to the best of our knowledge complete and accurate at the time of publication Rice Lake Weighing Systems reserv...

Страница 3: ...sion Measurement 4 2 7 Average Function 5 2 7 1 Average Procedure 5 2 7 2 Clearing Average 5 2 8 Peak Function 5 2 8 1 Clearing Peak 5 2 9 Safety Strap 5 2 10 Carry Case Option 5 3 0 Configuration 6 3...

Страница 4: ...na Clamp Tension Meter ii Visit our website www RiceLake com Rice Lake continually offers web based video training on a growing selection of product related topics at no cost Visit www ricelake com we...

Страница 5: ...nderstood Failure to follow the instructions or heed the warnings could result in injury or death Contact any Rice Lake Weighing Systems dealer for replacement manuals Failure to heed could result in...

Страница 6: ...Battery displays when approximately 10 of battery life remains and blinks when automatic shutdown is imminent 6 Primary display with six 0 7 18 mm numeric digits displays live wire rope tension and is...

Страница 7: ...n extended period of time 2 2 Product Dimensions Figure 2 1 Product Diagram 2 3 Power Press to turn on the Dyna Clamp The startup sequence displays all LCD segments followed briefly by the software ve...

Страница 8: ...y grease to the drive rod threads to maintain proper lubrication 1 Loosen the drive rod 2 Zero the Dyna Clamp Section 2 5 Figure 2 2 Dyna Clamp On Wire Rope 3 Position the Dyna Clamp on the wire rope...

Страница 9: ...Clearing Average Follow this procedure when the running average needs to be cleared 1 Position the Dyna Clamp on the wire rope and tighten slightly so a little positive tension is measured by the dev...

Страница 10: ...ls through the available wire rope types and the average and peak values in the secondary top display area Section 3 3 4 on page 7 ActFun performs action based on the active function that is displayed...

Страница 11: ...counting from to followed by the c cal value displaying in the secondary display area The test can be single stepped 1 Press F key set to then immediately press to stop the auto scroll 2 Press to scro...

Страница 12: ...gh setpoint value then the setpoint 2 LED turns on If the tension measurement is within the specified range then neither setpoint LED is on Use the following steps to set the Range setpoint values 1 P...

Страница 13: ...s in the secondary top display area 4 Press to scroll through the available filter settings 5 With desired filter displayed press to accept displays 6 Press to save and return to tension mode 3 7 Unit...

Страница 14: ...if it is in the setup menu The auto off parameter Section 3 4 on page 8 also conserves battery life by turning the unit off if no activity is sensed for a set time The standard mode is recommended for...

Страница 15: ...before load is lifted Zero the Dyna Clamp with no load attached lbf kgf kN units causing confusion Select proper units Requires recalibration Requires authorized service Display reading not stable Exc...

Страница 16: ...ls are recharged immediately after Low Battery starts to blink Do not allow the batteries to discharge completely it may compromise the recharge life of the battery NiMH Cells have a lower capacity th...

Страница 17: ...4 Bobbin Replacement 1 Remove the bolt securing the wire rope bobbin to the frame and discard old bobbin 2 Align new wire rope bobbin with hole in frame 3 Torque bolt to 30 ft lb 40 N m with blue Loc...

Страница 18: ...c digits Secondary Display top Eight 0 5 13 mm numeric digits Function Keys Programmable Keys Two user programmable F1 F2 TEST UNIT PRINT SCROLL ACTFUN Permanent Keys Two fixed POWER ZERO Setpoints Tw...

Страница 19: ......

Страница 20: ...anada Mexico 800 321 6703 International 715 234 9171 Europe 31 0 26 472 1319 Rice Lake Weighing Systems is an ISO 9001 registered company Rice Lake Weighing Systems Specifications subject to change wi...

Отзывы: