6. Set the COM Port to the port you are using; Baud Rate to 57600; Device to LPC2388; Interface to None 

(ISP); and Oscillator (MHz) to 12.0. 

7. Check 

Erase blocks used by Hex File

.

8. Using the 

Browse...

 button, locate the updated firmware file (.hex).

9. Check 

Verify after programming

 and click 

Start

.

The update will take several moments. The status is displayed at the bottom of the Flash Magic window.

Figure 1-2. Flash Magic settings

10. When the update is complete, the CW-90/90X will display 

Version Updated. Press Enter to Reset 

Configuration

.

11. Press the Power button to power-down the CW-90.

12. Remove JP3 and JP4 and turn power back on. The firmware update is complete.