5.0
Calibration
This section provides information regarding the calibration of the BenchPro Postal scale.
5.1 Span Calibration
The BenchPro allows for calibration with weight values other than max capacity.
displays the alternate calibration
weights for each model.
Calibration should only be performed using certified tests weights and performed by the local scale distributor.
Turn off Gravity Compensation, see
prior to performing a calibration using certified weights.
The default capacity (
) setting is in lb if using the lb scale, otherwise the default capacity setting is in kg if using
kg calibration weights, change the
to the appropriate scale capacity of the scale model. The model number of the
scale is located on the serial tag on the bottom of the scale.
1. See
displays.
2. Press
until
displays.
3. Press
to accept.
or
displays.
4. Press
to scroll to
.
5. Press
to accept.
displays.
6. Press
until
displays.
See
to set Gravity Compensation (
) values if required.
7. Select the appropriate scale capacity from
.
8. Press
to scroll to chosen capacity.
9. Press
to accept.
displays.
10. Press
until
displays.
11. Press
to accept.
displays.
12. With no weight on the weight platter, press
to accept. A six digit value displays.
This is the internal counts of the load cell at zero weight.
13. Press
to accept and perform a zero calibration.
displays.
14. Press
once.
displays.
15. Press
to accept. XXX lb or XXX kg displays.
Model Scale Capacity Alternate Calibration Weights
12 x 14
70 kg
20, 50, 70 lb
12 x 14
35 kg
10, 20, 30 kg
Table 5-1. Alternate Calibration Weights