350-10-8S Digital Stretcher Scale

20

Visit our website 

www.RiceLake.com

5.2 Communication Protocols

The scale has two communication protocols, escape and maintenance protocol.

5.2.1 Escape Protocol

An escape protocol is where the escape (0X1B or ASCII 27) is used to indicate that there is a command following. On the PC 

side there must be a listener created by the vendor that will interpret this protocol. This listener must also take care of all the 

issues regarding data integrity to make sure that the data that was sent and received is valid. 
Two examples include:

• Scale initiated communication
• PC initiated communication

The escape protocol commands table shows (below) what can be sent across communications lines.

ESC characters that will be used is shown below.

PC Initiated

ESC Value

Request current values/settings

R

Diagnostics

A

Send scale control messages

C

PC Initiated

ESC Value

Send single reading

R

Send diagnostic response

Table 5-2.  Escape Protocol Commands

Name

ESC 

Character

ESC Value with 

Parameters

Description

Reading

R

R

Tells PC the scale is sending a reading; immediately following this is the value that is sent 

Example: <ESC><R>ESC><W0200.0<ESC>Nm<ESC>E

Weight

W

Wnnn.n

The patient weight (

Example: W02000 means 200.0

). If scale is overloaded or under loaded,  

999.99 is returned

Height

H

Hnnn.n

Patient height

BMI

B

Bnn.n

Patient BMI

Units

N

Nc

Indicates the units the values have been taken (

m=metric, c=constitutional

).

End of Packet (EOP)

E

E

Indicates the end of the command has been reached.

Diagnostics 

(request)

A

Accc

A request for a diagnostic test on certain parts of the scale (like battery life, load cells).

Diagnostics 

(response)

Z

Zccc

The response of the diagnostics done on the scale; values include error codes to indicate an 

issue, or all zeros (Z000) to indicate the scale is performing properly

Control (set a value)

C

Cccc=c

Sets the value of the scale’s global settings 

Example: <ESC><CUOM=m><ESC><E sets the unit of measurement

Table 5-3.  ESC Characters

Name of Control

Identifier

Unit

Unit of Measure (metric or constitutional)

UOM

c (m or c)

Table 5-4.  Scale Global Values and Identifiers

Содержание 350-10-8S

Страница 1: ...350 10 8S Digital Stretcher Scale Software Version 11525 Technical Manual PN 213455 Rev A August 18 2022...

Страница 2: ...respective companies All information contained within this publication is to the best of our knowledge complete and accurate at the time of publication Rice Lake Weighing Systems reserves the right to...

Страница 3: ...raining or obtained by calling 715 234 9171 and asking for the training department Revision History This section tracks and describes the current and previous manual revisions for awareness of major u...

Страница 4: ...2 Weighing 9 3 3 Hold Release Function 9 3 4 Preset Tare 9 3 5 Toggle Tare 9 3 6 Using the Body Mass Index BMI Function 10 3 6 1 LB Mode 10 3 6 2 KG Mode 10 3 7 Troubleshooting 11 4 0 Configuration 12...

Страница 5: ...rt 15 of the FCC Rules These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment This equipment generates uses...

Страница 6: ...rt the scale while someone is on the scale Do not allow minors children or inexperienced persons to operate this scale Do not use in the presence of flammable materials Do not use this product if any...

Страница 7: ...parts are included and undamaged If parts were damaged in shipment notify Rice Lake Weighing Systems and the shipper immediately Parts contained in the shipping box include Scale Manual Indicator Part...

Страница 8: ...located in the parts box in the original shipping box Use the following steps to attach the feet to the scale 1 Tip the scale onto its side using the recessed handle openings Ensure the scale base is...

Страница 9: ...minimal clearance between the scale base and the floor without having the scale base contacting the floor Figure 2 5 Scale Base Clearance 5 It is important to ensure that the scale is completely level...

Страница 10: ...remove the tilt stand bracket from the indicator to gain access to the load cell connection Figure 2 7 Remove Tilt Stand 2 Remove the four back retaining screws to remove the back cover to the indicat...

Страница 11: ...e used for an extended period of time If the LO BAT indicator activates for accurate weighing replace the batteries or connect the scale to an AC power source as soon as possible 2 3 3 Power Connectio...

Страница 12: ...I Pressing the BMI key enables access to the BMI Body Mass Index mode defaults when scale is turned on The patient is gets on the scale weight stabilizes and press the BMI key The display then asks fo...

Страница 13: ...y the patient 1 Press to turn on the scale 0 0 appears on the display along with the ZERO annunciator 2 Place additional item s on the scale 3 Press until the display returns to 0 0 and NET annunciato...

Страница 14: ...adjust the height value 5 Press 6 The BMI value and BMI annunciator are shown on the display Press to return to the Weighing mode and the BMI function will be turned off 3 6 2 KG Mode 1 Ensure that t...

Страница 15: ...high E07 AD too low E10 Overload Scale has been overloaded Remove load from scale E4L BAT Battery low but still usable one bar left on indicator display E4U Battery low and unstable no bars left on in...

Страница 16: ...the menus and or parameters Press to enter a displayed menu and or parameter Press or to scroll through values Press to save the displayed selection and move to the next parameter 4 2 2 Enter Numbers...

Страница 17: ...cated in the back of the scale to enter the setup parameters for the scale Use a small paper clip small screwdriver or other similar object to press the setup switch 4 Once the setup switch is pressed...

Страница 18: ...Use the Up Down arrow keys to increment decrement numbers 7 When done press ENTER key to move to the next parameter LOAD LOAD This is the amount of weight applied during calibration Can also be change...

Страница 19: ...10 20 50 100 Values in lb 1 2 5 10 20 50 100 200 0 2 0 5 1 0 2 0 5 0 10 0 20 0 0 1 The display toggles between a numeric value and ROUND The decimal point location is set to the DISP parameter display...

Страница 20: ...use the following steps 1 Press the ENTER key to change values 2 Use the Up Down arrow keys to change the available parameters 3 When done press ENTER key to move to the next parameter UNITS UNITS Un...

Страница 21: ...e scale to enter the setup parameters for the scale Use a small paper clip small screwdriver or other similar object to press the setup switch on the back of the indicator under the cover 4 Prog displ...

Страница 22: ...r on the scale until ID flashes 2 Press again 3 Continue to press to scroll through the various menu items 4 Once complete press again then Done displays 5 Press to start the weighing process Figure 4...

Страница 23: ...weight was either in motion or out of range If displaying weight and not BMI the scale will send out the following 21 character string xxxxxxxxx SP uu SP mmmmm SP CR LF Example 60 1 KG PATIENT SP WEI...

Страница 24: ...pe Protocol Commands Name ESC Character ESC Value with Parameters Description Reading R R Tells PC the scale is sending a reading immediately following this is the value that is sent Example ESC R ESC...

Страница 25: ...Zero the scale F Show flash values used for the first flash process L USB On Off not available on USB communication Table 5 5 Maintenance Protocol Commands Escape Protocol Example 1 Escape Protocol E...

Страница 26: ...ing the use of the USB driver Driver protections may need to be temporarily disabled on Windows 10 or later computers to allow for the installation of the USB driver 1 Connect the scale s indicator to...

Страница 27: ...verall appearance of the entire scale for any obvious signs of damage Inspect the condition of the AC power adapter cord for cracking fraying or for broken or bent prongs 6 2 Cleaning Proper care and...

Страница 28: ...s Automatic power off can be configured Environmental Operating Temperature 50 F to 104 F 10 C to 40 C Storage Temperature 0 F to 122 F 10 C to 50 C Humidity 85 Capacity 1000 lb x 0 2 lb 450 kg x 0 1...

Страница 29: ......

Страница 30: ...54868 USA U S 800 472 6703 Canada Mexico 800 321 6703 International 715 234 9171 Europe 31 0 26 472 1319 Rice Lake Weighing Systems Specifications subject to change without notice www ricelake com PN...

Отзывы: