background image

Configuration

21

5.4.2

Programming Mode

The Programming Mode allows setup of the indicator to determine how the scale reads the weight and displays output.

Figure 5-5. Programming Menu Layout

Parameter

Description

FULL

Full Capacity – enter value of the scale capacity; default: 2500 kg

LOAD

Load – weight applied during calibration; default: 160.0 kg  NOTE: can also be changed in the calibration menu

G-CAL

........  default: 00000

G-LOC

...............  default: 00000

ASTART

Auto-Start – weight process start limit; maximum (full capacity)/10; determines when the weight algorithm starts (when the 

----

 is dis-

played), below this value the scale displays live weight; default: 2.0 kg

ARW

Auto Reweigh – restarts the weight algorithm if the weight changed by more than this value; default: 2.0 kg

SAL

Semi Auto Live – interval between weight displays during the algorithm process; default: 0.5

ROUND

Scale Resolution – display toggles between a numeric value and 

ROUND

; decimal point location is set to the DISP parameter; 

selections: 0.1 kg (default), 0.5, 1, 2, 5, 10, 20, 0.2

DISP

Decimal Point – display toggles between a numeric value and 

DISP

default: 00000

BAUD

Baud Rate – set required baud rate; selections96 (default), 48, 1152, 576, 384, 288, 192, 144; end zeros are not displayed

ATOL

Algorithm Initial Tolerance – display toggles between a numeric value and 

ATOL

default: 10; 

value above 255 (max) does not allow indicator to proceed and returns to the previous value

ALEN

Algorithm Initial Exponent – display toggles between a numeric value and 

ALEN

default: 8;

Value above 10 (max) does not allow indicator to proceed and returns to the previous value.

ATOUT

Algorithm Maximal Exponent – display toggles between a numeric value and 

DISP

default: 10;

Values above 15 (max), does not allow indicator to proceed and returns to the previous value.

TOFF

Auto Off Timer – turns unit after the set number of minutes; 0-9 minutes; 0 = always on; default: 5 min
battery power only, does not work if using an external power supply

SPACE

Space – number of new lines after print; default: 7

UNITS

Units – select the unit of measure; selections: LB, KG,  KG/LB (default)
NOTE: When indicator is set to LB or KG, toggling between them is not available in the weighing mode.

Table 5-3.  Program Mode Parameters 

PROG

FULL

LOAD

G-CAL

G-LOC

ASTRT

ARW

SAL

ROUND

DISP

value

value

value

value

value

value

value

0.2

1.0

0.5

2.0

10.0

5.0

0.1

00000

0.00

0.0

0.000

0.0000

BAUD

ATOL

ALEN

ATOUT

TOFF

SPACE

UNITS

OP

SAVE

value

value

value

value

P00-0

KG & LB

KG

5

3

4

2

1

0

8

9

6

7

96

1152

48

576

384

288

144

192

LB

Содержание 160-10-7

Страница 1: ...160 10 7 Low Profile Digital Physician Scale Firmware Version 11525 Technical Manual PN 185558 Rev C August 18 2017...

Страница 2: ...marks of their respective companies All information contained within this publication is to the best of our knowledge complete and accurate at the time of publication Rice Lake Weighing Systems reserv...

Страница 3: ...ration 7 3 1 Powering the Scale 7 3 2 Set Units 7 3 3 Weigh Mode 8 3 4 Hold Release 8 3 5 Preset Tare 8 3 6 Toggle Tare 8 3 7 Body Mass Index 8 3 7 1 LB Mode 8 3 7 2 KG Mode 9 4 0 Communications 10 4...

Страница 4: ...ke com Rice Lake continually offers web based video training on a growing selection of product related topics at no cost Visit www ricelake com webinars 6 1 2 Cleaning 24 6 2 Troubleshooting 24 6 3 Sp...

Страница 5: ...s can be viewed or downloaded from the Rice Lake Weighing Systems website at www ricelake com manuals Figure 1 1 Rice Lake Low Profile Digital Physician Scale RL160 10 7 1 1 Standard Features Capacity...

Страница 6: ...lt in serious injury or death Subjects with disabilities or who are physically frail should always be assisted on and off the scale Do not use the scale on slippery surfaces such as a wet floor Do not...

Страница 7: ...160 10 7 must be properly packed with sufficient packing materials prior to being moved or shipped Retain the original shipping materials and container if possible Damage caused by improper packaging...

Страница 8: ...nd remove the tilt stand bracket from the indicator Figure 2 2 Remove Tilt Stand 2 Remove the four back retaining screws to remove the back cover to the indicator Figure 2 3 Remove Back Cover 3 Plug t...

Страница 9: ...secure it Batteries must be removed from the unit prior to long periods of disuse or long term storage 2 3 3 Optional Power Connections Rice Lake Weighing Systems offers optional AC adapters to use wh...

Страница 10: ...160 10 7 Low Profile Digital Physician Scale 6 Visit our website www RiceLake com 2 3 4 Dimensions Figure 2 7 Indicator Dimensions 6 3 160 mm 6 5 160 mm 2 9 72 4 mm...

Страница 11: ...ired by subject from the scale Example oxygen unit or other equipment 3 CLEAR return to weigh mode after BMI value displays 4 Print long key press sends data out from the RS 232 port or USB port LB KG...

Страница 12: ...subject step on the scale with accessory items The net weight displays 5 The weight of accessory items remain stored in memory To cancel the tare weight press and hold until the weight returns to 0 0...

Страница 13: ...h the weight at 0 0 and the ZERO annunciator on step onto the scale The weight value displays the LOCK annunciator is on and an audible tone indicates the end of the weighing process 2 Press The BMI a...

Страница 14: ...r five seconds release the button as the weight was either in motion or out of range In the weigh mode the scale sends the 21 character string xxxxxxxxx SP uu SP mmmmm SP CR LF Where xxxxxxxxx is the...

Страница 15: ...tes the unit of the values m metric c constitutional End of Packet EOP E E Indicates the end of the command Diagnostics request A Accc A request for a diagnostic test on parts of the scale e g battery...

Страница 16: ...ed Basic information on USB driver installation using Windows is described in the following steps and serves only as an example A USB driver can be downloaded from the Rice Lake Weighing Systems websi...

Страница 17: ...Download Page 3 Locate the Titanium USB Driver Click on Download to open and download the driver to a local computer Figure 4 6 Example of Drivers List Page 4 When the USB cable is connected to the in...

Страница 18: ...ed in the device manager of the computer Figure 4 8 Device Manager 9 To configure a printer using the USB driver open the software driver Parpar in the device manager Figure 4 8 The port assigned to t...

Страница 19: ...ter PATIENT WEIGHT one space between the weight value and units lb and seven CR LF after Figure 4 10 IT Example IT Information Only GROSS WEIGHT 199 8 lb TARE WEIGHT 0 0 lb PATIENT HEIGHT 6ft 00 0in P...

Страница 20: ...of software packages available there are many ways to pair and connect to the scale The steps used to pair the scale will vary depending on the device software being used The scale is set up to be dis...

Страница 21: ...window right click on the device icon and select Properties 14 Select the Hardware tab to check the correct Comm and then click OK to confirm that Figure 4 14 Hardware Tab 15 Choose any communication...

Страница 22: ...Press or to increment decrement numbers Press to move to the next digit Press to save value and move to the next parameter 5 1 3 Save and Return to Main Menu When a parameters selection value is corre...

Страница 23: ...alue flashes 3 Press TEST displays 4 Press to scroll to USER 5 Press BAUD and it s value flashes Parameter Description BAUD Baud Rate set required baud rate selections 96 default 48 1152 576 384 288 1...

Страница 24: ...removing the back cover Figure 5 3 Back of Indicator Use the following steps to enter into setup mode 1 Ensure the scale is off 2 Remove the back plate by loosening the four screws as shown in Figure...

Страница 25: ...ns 96 default 48 1152 576 384 288 192 144 end zeros are not displayed ATOL Algorithm Initial Tolerance display toggles between a numeric value and ATOL default 10 value above 255 max does not allow in...

Страница 26: ...IB 3 Press A numeric value displays weight to be used for calibration 4 Only calibrate in LB 5 Press The digit on the right flashes Press or to change the value of digit Press to move to the next digi...

Страница 27: ...ress displays 7 When PUT and value flashes on display put the weight on the platform 8 Press and then Save displays 9 Press DONE displays 10 Press three times to the top level DONE parameter 11 Press...

Страница 28: ...rect failures prior contacting service personnel for assistance Symptom or Error Possible Cause Corrective Action Scale does not turn on Dead battery Replace batteries or connect scale to a power sour...

Страница 29: ...ttery Power Rechargeable Class III or 6 X 1 5AA Non Class III Battery Life 25 hours continuous automatic power off can be configured Data Communications Connections RS 232 with RJ 45 jack USB Baud Rat...

Страница 30: ...actually exists and was not caused by accident misuse neglect alteration improper installation improper repair or improper testing RLWS shall be the sole judge of all alleged non conformities Such equ...

Страница 31: ......

Страница 32: ...nada Mexico 800 321 6703 International 715 234 9171 Europe 31 0 26 472 1319 Rice Lake Weighing Systems is an ISO 9001 registered company Rice Lake Weighing Systems Speci cations subject to change with...

Отзывы: