Configuration
21
5.4.2
Programming Mode
The Programming Mode allows setup of the indicator to determine how the scale reads the weight and displays output.
Figure 5-5. Programming Menu Layout
Parameter
Description
FULL
Full Capacity – enter value of the scale capacity; default: 2500 kg
LOAD
Load – weight applied during calibration; default: 160.0 kg NOTE: can also be changed in the calibration menu
G-CAL
........ default: 00000
G-LOC
............... default: 00000
ASTART
Auto-Start – weight process start limit; maximum (full capacity)/10; determines when the weight algorithm starts (when the
----
is dis-
played), below this value the scale displays live weight; default: 2.0 kg
ARW
Auto Reweigh – restarts the weight algorithm if the weight changed by more than this value; default: 2.0 kg
SAL
Semi Auto Live – interval between weight displays during the algorithm process; default: 0.5
ROUND
Scale Resolution – display toggles between a numeric value and
ROUND
; decimal point location is set to the DISP parameter;
selections: 0.1 kg (default), 0.5, 1, 2, 5, 10, 20, 0.2
DISP
Decimal Point – display toggles between a numeric value and
DISP
; default: 00000
BAUD
Baud Rate – set required baud rate; selections: 96 (default), 48, 1152, 576, 384, 288, 192, 144; end zeros are not displayed
ATOL
Algorithm Initial Tolerance – display toggles between a numeric value and
ATOL
; default: 10;
value above 255 (max) does not allow indicator to proceed and returns to the previous value
ALEN
Algorithm Initial Exponent – display toggles between a numeric value and
ALEN
; default: 8;
Value above 10 (max) does not allow indicator to proceed and returns to the previous value.
ATOUT
Algorithm Maximal Exponent – display toggles between a numeric value and
DISP
; default: 10;
Values above 15 (max), does not allow indicator to proceed and returns to the previous value.
TOFF
Auto Off Timer – turns unit after the set number of minutes; 0-9 minutes; 0 = always on; default: 5 min
battery power only, does not work if using an external power supply
SPACE
Space – number of new lines after print; default: 7
UNITS
Units – select the unit of measure; selections: LB, KG, KG/LB (default)
NOTE: When indicator is set to LB or KG, toggling between them is not available in the weighing mode.
Table 5-3. Program Mode Parameters
PROG
FULL
LOAD
G-CAL
G-LOC
ASTRT
ARW
SAL
ROUND
DISP
value
value
value
value
value
value
value
0.2
1.0
0.5
2.0
10.0
5.0
0.1
00000
0.00
0.0
0.000
0.0000
BAUD
ATOL
ALEN
ATOUT
TOFF
SPACE
UNITS
OP
SAVE
value
value
value
value
P00-0
KG & LB
KG
5
3
4
2
1
0
8
9
6
7
96
1152
48
576
384
288
144
192
LB