Simple Auto Calibration
:
Press the key about 3 seconds. When the display is flashing and
showing “Calibration Weight Value”, please accord with the weight to put the
same weight mass. When the display is stable and stop flashing, please take
away the mass. Auto calibration is finished.
Remark:
The “Calibration Weight Value” shown on the display is normally based on the
latest calibration weight value. Calibration weight value can be changed as
following:
Turn on the scale. During the scale in self-testing, press the key and the
“func” key and then the display shows “CC” and enter to the function selections
mode press the key to enter into
calibration mode. Press the key to change the figure. Press the
key to move the digit and confirm the key in value.
When the display is flashing and showing the weight value, put the mass on the
scale. When the display is stable and stop flashing, please take away the mass.
Auto calibration is finished.
Remark:
Mass value:
Any over (full calibration x 10%) weight of key in value is acceptable to calibrate.
Suggestion:
Generally, if the calibrated weight is over half load capacity or full load capacity, it
will be good for getting the precise accuracy.
Σ
+