Bench Scale 940C • Operating Manual and Adjustment Guide

2.7 Adjustment

Owing to the high resolution of the scale, the different gravitational force (g value) at different plac-
es of use must be taken into account.
To this end, the scale should be adjusted using test weights (accuracy class at least M1, preferably
F2) before being taken into operation.

Adjustment should be performed only when necessary.

Inaccurate values on the display or replacement of the load cell or the weighing electronics can
necessitate an adjustment.

2.7.1 Conditions

!

Switch on the scale and wait 15 minutes before performing the adjustment.

Adjustment can be performed only with fixed weight values. Weights corresponding to half of
and the full maximum safe load are required.

2.7.2 Adjustment procedure

Align the scale exactly using the integral bubble level (spirit level).

Remove any load from the scale.

Power off the scale by pressing the 

 key.

Press and hold down the 

 key and power on the scale using the 

 key.

Hold down the 

 key until such time as the internal measured value appears on the display

(approx. 2s).
(e.g. 



With the weighing platform at no load, press the 

 key.

The first adjustment point (zero point) is saved.



 appears briefly on the display.

The weight value for the second adjustment point (half the maximum safe load of the scale)
appears on the display.

Subject the scale to the weight indicated and confirm by pressing the 

 key.



 appears briefly on the display.

The weight value for the third adjustment point (maximum safe load of the scale) appears on
the display.

Subject the scale to the weight indicated and confirm by pressing the 

 key.



appears briefly on the display and then 



.

End adjustment by pressing the 

 key.

Power off the scale by pressing the 

 key.

Содержание 940C/10

Страница 1: ...Bench Scale 940C Operating Manual Adjustment Guide RHEWA WAAGENFABRIK August Freudewald GmbH Co KG ...

Страница 2: ...s documentation cannot be taken to mean that such names are not protected by trade mark and brand protection legislation and therefore open to use by anyone even if this is not expressly stated Technical modifications Rapidly advancing technological developments and shorter product cycles prevent us from bringing this docu mentation exactly into line with the appliance functions and properties Sho...

Страница 3: ...ce 5 2 4 Startup 6 2 4 1 Inserting the batteries 6 2 4 2 Aligning the scale 6 2 5 Battery operation 6 2 6 Keypad 7 2 6 1 Power on off 7 2 6 2 Zero 7 2 6 3 Tare 7 2 6 4 Gross display 7 2 6 5 Putting mixtures together 7 2 6 6 Discharge weighing 7 2 7 Adjustment 8 2 7 1 Conditions 8 2 7 2 Adjustment procedure 8 Specifications Chapter 3 9 3 1 Error messages 9 3 2 Declaration of conformity 10 ...

Страница 4: ...4 925 Tabletop Scale Operating Manual and Adjustment Guide ...

Страница 5: ...significant temperature fluctuations e g direct sunlight outlets of air conditioning or heating systems No vibrations No magnetic fields Do not place in the immediate vicinity of transmitting devices mobile phones radio equipment etc 2 3 Care and maintenance This scale is a precision instrument for which great importance was attached to exact meas urement results during the manufacturing and setti...

Страница 6: ...e is a bubble level spirit level next to the display Using the adjustable feet align the scale so that the air bubble in the level is in the middle of the circle The scale must sit firmly on all four feet Attach the weighing platform lengthwise or crosswise as required 2 5 Battery operation New batteries have a service life of approx 500 hours To achieve a long battery life power off the scale whe...

Страница 7: ...weighing platform The scale displays the weight of the empty container with a preceding minus sign On pressing the key again appears on the display The tare has been cleared 2 6 4 Gross display With a tared scale press the key The display toggles between GROSS and NET each time the key is pressed 2 6 5 Putting mixtures together Press the key again each time a component is added 2 6 6 Discharge wei...

Страница 8: ...ment procedure Align the scale exactly using the integral bubble level spirit level Remove any load from the scale Power off the scale by pressing the key Press and hold down the key and power on the scale using the key Hold down the key until such time as the internal measured value appears on the display approx 2s e g With the weighing platform at no load press the key The first adjustment point...

Страница 9: ...pply 4 LR14 batteries 1 5 V Mains option using 230V 50Hz power adapter available as an accessory Power consumption 0 3 W Net weight Approx 1 5 kg without batteries Message Explanation Remedy Cannot power up Batteries too weak Replace batteries Indication is unstable A stable indication is not obtained Wait for a stable indication Keep away from draughts and vibrations Renew batteries AD converter ...

Страница 10: ...ponds to the requirements of the following EC directives correspond aux exigences des directives CE suivantes 2004 108 EC Applied standards EN 61000 6 1 2007 EN 61000 6 3 2007 EN 61000 3 2 2006 EN 61000 3 3 1995 A1 2001 A2 2005 RHEWA WAAGENFABRIK A Freudewald GmbH Co Hersteller Manufacturer Fabricant RHEWA WAAGENFABRIK August Freudewald GmbH Co KG Postfach 10 01 29 D 40801 Mettmann Typ Type Type 9...

Отзывы: