background image

Operating Instructions

Type 70 High Pressure Switching Valves
Model 7010 Sample Injector

1.0 DESCRIPTION

Rheodyne’s Type 70 valves are high

pressure switching valves: 7000, 7000L,
7030, 7030L, 7040, 7040L, 7060 and 7060L.
Unless otherwise indicated, statements about
non-L models apply also to the L models. 

Model 7010 is a sample injector. An

accessory Loop Filler Port (P/N 7012) is
needed to load sample into Model 7010. If
your valve is to be operated with a Rheodyne
actuator, refer also to the separate actuator
instructions.

Figure 1 shows a schematic flow diagram

of each of the switching valves. The circles
represent the ports in the valve stator. The
dark grooves are the connecting passages in
the rotor seal.

Rotation of the valve shaft through 60°

switches the valve from one position to the
other. Models 7010, 7000, 7030, and 7040
are two-position valves with stops that
restrict the rotation of the valve to 60° turns.
Model 7060 is a six-position valve. A spring-
loaded detent mechanism is included which
allows continuous rotation and insures that
the shaft “falls into” each of the six positions
at the precise 60° spacing.

2.0 SUPPLIED WITH THE VALVE

Supplied with the valve in a separate bag

are RheFlex

®

stainless steel fittings sets for

all ports and the following items. A 20 µL
sample loop is supplied with Model 7010. 
A joining tube is supplied with Model 7040.

• Hex Key(s)
• Mounting Screws (2)

3.0 SPECIFICATIONS

• Maximum Operating Pressure: Model

7010 and Type 70 non-L models - 
48 MPa (483 bar, 7000 psi); L models -
34 MPa (345 bar, 5000 psi)

• Maximum Operating Temperature:

Models 7010, 7000, 7030, 7040, and 
L models - 150°C; Model 7060 - 80°C

• Flow Passage Diameters: Models 7010,

7000, 7030, 7040 - 0.5 mm (0.018") and
0.6 mm (0.024"); Model 7060 - 0.5 mm
(0.018") and 0.4 mm (0.016"); 
L models - 1 mm (0.040")

• Wetted Surfaces: stainless steel and an

inert polymer

• Ports in all models accept 10-32 male

threaded fittings

4.0 IMPORTANT SAFETY NOTICES
4.1 Warning: 
(Using 7010 as an injector):
When using sample loops larger than 
100 µL, shield yourself from mobile phase
coming out of the needle port when the valve
is turned from INJECT to LOAD. Example:
1 mL loop ejects 20 µL upon decompression
from 19 MPa (200 bar, 2898 psi).
4.2 Caution: Do not mount the valve with
the ports facing up. Leakage due to a
damaged rotor seal or loose fittings can
cause the bearings to corrode.
4.3 Caution: Rinse the valve thoroughly
after using buffer solutions to prevent salt
crystals from forming, which can cause
damage to the rotor seal and stator face
assembly.

5.0 INSTALLATION

All valves can be panel mounted:
a) Remove the handle assembly by

loosening the knob set screw(s). 

b) Fasten the valve to the panel using the

two mounting screws supplied.

On Model 7060, the two screws that hold

the detent mechanism on the valve should
be removed before fastening the valve to the
panel. If the detent mechanism needs to be
disabled, as is required when installing a
pneumatic actuator around Model 7060,
refer to Section 8.4.

5.1 MODEL 7010 INJECTOR

Model 7010, as a sample injector, connects

to different system components.

a) Connect an accessory injection port to

Port 5 and the vent line to Port 6. Place the
injection port at the same horizontal level as
Vent Line 6 to avoid siphoning.

b) Connect the pump to Port 2 and the

column to Port 3. Leave the column
disconnected from the valve during initial

Position A

Position B

Models 7000 & 7010

Two-Position, Six-Port

Position A

Position B

Model 7030

Three-Way

Position B

Position A

3

2

4

5

1

3

2

4

6

5

1

6

Model 7060

Six-Position

3

2

4

6

7

5

1

Position A

Position B

Model 7040

Four-Way

2

4

6

5

1

3

2

4

6

5

1

3

Fig. 1. Flow diagrams of Rheodyne Type 70 high pressure switching valves.

Page 1 of 4

2320666B

6/02

2

4

6

5

1

3

2

4

6

5

1

3

Position A

Position B

Rheodyne LLC • A Unit of IDEX Corporation • 600 Park Court, Rohnert Park, CA 94928 • (707) 588-2000 • Fax (707) 588-2020

www.rheodyne.com

Содержание Model 7010

Страница 1: ... in all models accept 10 32 male threaded fittings 4 0 IMPORTANT SAFETY NOTICES 4 1 Warning Using 7010 as an injector When using sample loops larger than 100 µL shield yourself from mobile phase coming out of the needle port when the valve is turned from INJECT to LOAD Example 1 mL loop ejects 20 µL upon decompression from 19 MPa 200 bar 2898 psi 4 2 Caution Do not mount the valve with the ports f...

Страница 2: ... 4 and 6 and proceed as follows a Mount the new isolation seal onto the shaft with the open side facing the handle b Be sure that the rotor seal is correctly oriented as shown in Figure 6 with rotor seal grooves facing the stator and with the notch in the metal rim of the rotor seal in line as shown the notch also faces the stator c In replacing the stator ring be sure that the two stop pins are s...

Страница 3: ...good wear resistance However Vespel is susceptible to alkaline attack deteriorating rapidly when used with solutions of pH over 10 PEEK and Tefzel1 rotor seals are available for alkaline applications 9 3 ACCURACY OF SAMPLE LOOPS Sample loop sizes are not actual values The actual volume can differ by 10 for a 20 µL loop and 5 for a 10 mL loop There is a greater difference for smaller loops Detent H...

Страница 4: ... the warranty This limited warranty is Rheodyne s sole warranty of its products and all other warranties of merchantability or fitness for any particular purpose are hereby disclaimed Under no circumstances will Rheodyne be liable for any consequential or incidental damages attributable to a claimed failure of a Rheodyne product even if Rheodyne has been placed on notice of possibility of such dam...

Отзывы: