Testing and Care
Page 20 of 30
Mother Transmitters User Guide
0510-1020-O
3. Verify that the warranty expiration date
stamped on the transmitter has not
expired. If the warranty period has
expired, discard and replace the
transmitter immediately.
4. Your facility must keep records of test
and transmitter inspection.
Transmitter
Clearing
Device
1. Verify that the warranty expiration date
that is stamped on the device is not
expired. If the warranty period has
expired, discard and replace the device
immediately.
2. Visually inspect the device for damage
or loose parts.
Test the operation of the transmitter clearing
device using the Transmitter Tester (PN
9450-0050). The transmitter tester will
detect whether or not a transmitter is
emitting a signal, but cannot indicate the
strength of the signal
1. Place the transmitter tester directly on
the transmitter.
2. Press and hold the button on the left
side of the transmitter tester.
3. The device beeps once when you
initially press the button.
4. While holding the button in, the indicator
light flashes and a tone sounds once
per second
5. Wait for at least 3 flashes of the
indicator light and 3 tones from the
transmitter tester to verify that the
device is functioning correctly.
6. If the device fails this test, check the
battery in the tester. The transmitter
tester requires a 9V battery.
Содержание Safe Place Mother
Страница 1: ...Mother Transmitters User Guide ...
Страница 26: ...Testing and Care Page 24 of 30 Mother Transmitters User Guide 0510 1020 O This page intentionally left blank ...
Страница 30: ...Specifications Page 28 of 30 Mother Transmitters User Guide 0510 1020 O This page intentionally left blank ...