Do not immerse the device in water.

Do not clean with chemicals.

No part of REVITIVE Ultrasound is suitable for dishwasher use.

Instructions for use

Troubleshooting

Important:

 It is recommended to clean the product after every use.

Clean the Treatment Head with a tissue or a soft, damp cloth. This will remove 
any excess Ultrasound Gel.

STORAGE

CLEANING RECOMMENDATIONS

REVITIVE Ultrasound:
When not in use, store the device with the adaptor in a dry room and protect 
it against extreme moisture, heat and direct sunlight.

Air pressure, humidity and temperature limits for storage:

REVITIVE Ultrasound Gel: 
Wipe nozzle clean and re-cap after each use. Store in a cool, dry place.

-10 °C

50 °C

20 %

93 %

1060 hPa

700 hPa

700 hPa

1060 hPa

No vibration

Ultrasound waves, with a vibra-
tion of 1-million times per sec-
ond are too fast to see or feel. 
Therefore, you will not feel any 
sensation when using REVITIVE 
Ultrasound.

To confirm your REVITIVE 
Ultrasound is operating, 
follow the treatment head 
test (See page 14 for 
details).

The temperature safety switch 
has tripped and the unit has 
switched off.

Allow the unit to cool down.

No vibration or 

indicators

The product has failed.

Contact your dealer.

Unit will not turn on

Temperature is too high.

Allow to cool.

Time indicator 

flashing

Poor contact between the 
treatment head and the body, 
(not enough pressure is being 
applied).

Apply a generous amount 
of Ultrasound Gel to create 
good contact between 
the Treatment Head and 
the body.  Treatment is 
automatically restored when 
good contact is restored.

Problem                 Possible Cause                      Solution

3269_IFU02_14269060.indd   16-17

24/07/2017   11:59

Содержание UT1033

Страница 1: ...r UT1033 PLEASE READ THE USER S MANUAL CAREFULLY BEFORE USING THIS PRODUCT USE ONLY AS DIRECTED If symptoms persist consult your healthcare professional User s Manual 3269_IFU02_14269060 indd 1 24 07...

Страница 2: ...gs and Cautions 8 Instructions for use 9 16 How to use REVITIVE Ultrasound 9 11 Therapy Application Table 12 13 Treatment Head Test 14 15 How to use the Ultrasound Gel 15 Cleaning Recommendations 16 S...

Страница 3: ...opening the carton please check that the following components are provided If you think anything is missing please contact us using the helpline numbers on the back of this booklet What is inside the...

Страница 4: ...ll or reproductive organs Used over open wounds or fragile or damaged skin eg eczema Used on the front of the neck over the carotid sinus Used over spinal abnormalities e g spina bifida following lami...

Страница 5: ...and consult your health care professional The socket outlet should be installed near the equipment and should be easily accessible Use only the accessories supplied by or purchased from the manufactu...

Страница 6: ...e Time Indicators When in use the chosen Time Indicator will be on constantly and the treatment will count down for the chosen Time duration If the Time Indicator is flashing there is not enough press...

Страница 7: ...inutes per treatment head area 1 0 4 LOW 4 minutes per treatment head area 2 0 8 LOW 4 5 minutes per treatment head area 3 1 2 MEDIUM 2 minutes per treatment head area 4 1 6 MEDIUM 2 5 minutes per tre...

Страница 8: ...ater will visibly vibrate demonstrating that the Treatment Head is working correctly Important The Treatment Head only vibrates if the conductive fluid water or gel covers the entire surface of the Tr...

Страница 9: ...Pa 700 hPa 1060 hPa No vibration Ultrasound waves with a vibra tion of 1 million times per sec ond are too fast to see or feel Therefore you will not feel any sensation when using REVITIVE Ultrasound...

Страница 10: ...rical Device Type BF IEC 60601 1 6 1l Serial number is on the underside of the device and on the packaging Directive 93 42EEC Annex 1 13 3 d EN 15223 1 2012 Lot Number including year YY and Month MM o...

Страница 11: ...ICATIONS EQUIPMENT AND EQUIPMENT OR SYSTEMS THAT ARE NOT LIFE SUPPORTING Technical Specifications Technical Specifications REVITIVE Ultrasound is intended for use in the electromagnetic environment sp...

Страница 12: ...l be calculated from the original purchase date Exclusions Actegy Ltd shall not be liable to replace or repair the goods under the terms of the guarantee where Damage or the fault is due to improper u...

Страница 13: ...management standard for medical devices Made in PRC Shanghai International Holding Corp GmbH Europe Eiffestra e 80 20537 Hamburg Germany Shenzhen Dongdixin Technology Co Ltd No3 Building Xilibaimang...

Отзывы: