SERENA Tecnico GB R-06 del 10.05.2019
00/ Per Bertani/ Manuale Tecnico Serena/ Serena
11
1.7 VACUUM DRY
CALIBRATION
INCREASE BAR
TIMED REFILLING
VACUUM MAX
VACUUM DRY
The values (D) indicate the range in which the bacteriological filter is activated
VACUUM DRY
MIN 0600 mbar
MAX 0700 mbar
1.8 VAPORIZER TEMPERATURE
TIMED REFILLING
VACUUM MAX
VACUUM DRY
VAPORIZER
TEMPERATURE
The following screen will appear:
The values indicate the range in which occurs the injection of water that becomes steam.
VAPORIZER
TEMPERATURE
MIN 150 °C
MAX 160 °C