65
Fig. 37: C
ycle editor
Element
Function
NB1 Main menu
To return to the main menu.
NB6 Cancel
Cancels editing of the cycle.
PE
Parameter settings
Displays the parameter configured for the active cycle.
DZ
Total duration of the cycle
Displays the total duration of the cycle (the total duration
consists of the two parameter sets (A/B) and the
repeats).
The total duration of a cycle is limited to 99 hours.
GS5 Delete
Deletes the parameters of the cycle.
GS6 Save
Saves the cycle.
Press
in cycle mode to select the cycle editor and to edit the activated cycle.
The total duration of a cycle may not exceed 99 hours. A total duration of more than 99
hours cannot be saved and will be shown in red.
The process can be cancelled by pressing the
button.
All settings made will then be discarded.
Parameters can only be changed in cycle mode if editing has been activated using the
button.
Press on the parameter you wish to edit.
Turn the dial until the desired value is displayed.
Press on the parameter again or select a different parameter to accept the set value.
NB1
NB6
GS5
GS6
PE
DZ