26.10.1998
Retsch GmbH
9
Doc. No .E98.052.9999
Requirements for the installation site
Ambient temperature:
5°C to 40°C
When the ambient temperature exceeds or falls below that
specified, the electronic and mechanical components may
be damaged, and performance data changed to an unknown
extent.
Humidity:
Maximum relative humidity 80% at temperatures up to 31°C;
linear decline down to 50% relative humidity at 40°C.
At higher humidity, the electronic and mechanical
components may be damaged, and performance data
changed to an unknown extent.
Installation site
–
altitude
max. 2000 m above mean sea level
Fig. 1
Installation
Set up the BB 100 only on a solid, level surface. The jaw crusher
should be anchored to provide sufficient protection against tipping;
use the holes in the feet of the BB 100 for this purpose; see
Fig. 1
.
The vibrations transmitted to the environment by the free mass
moments in the BB 100 will be only barely perceptible.
Electrical connection
Refer to the data plate for information on the voltage and
frequency at which the BB 100 is to be operated.
Ensure that the values shown there correspond to those for the
local power supply.
Use the supplied power cord to connect the BB 100 to the
power supply.
When connecting the power cord to the power source, be sure
that the supply circuit is fused in accordance with local codes.
!
The BB 100 will not exhibit the degree of protection against
tipping prescribed in EN 61010 unless it is anchored to the
floor.
Components could be damaged.
Hazard of crushing arms and feet.
Failure to observe the values on the data plate can cause
damage to the electrical or the mechanical components or
both.