Packaging, Transport and Installation
20
4.5 Electrical Connection
WARNING
W4.0005
Danger to life due to electric shock or fire
Incorrect connection to the power supply may result in parts of the housing or
cables being live and in fires starting.
Serious injuries or death due to an electric shock.
Serious injuries or death due to fires.
•
The device may only be connected by a qualified electrician.
NOTICE
N9.0022
Electrical connection
Failure to observe the values on the type plate
Electronic and mechanical components may be damaged.
•
Connect the device only to a mains supply matching the values on
the type plate.
Information about the required voltage and frequency of the device can be found on the
type plate .
The device may only be connected to the power supply using the connection cable
supplied.
The circuit breaker for connecting the power cable to the power supply at the installation
site should be suitable for higher start-up current and correspond to a type C
characteristic (slow blow fuse).
NOTICE
N10.0005
Electrical connection
Wrong direction of rotation on drive motor
Electronic and mechanical components may be damaged.
Insufficient grinding of the sample material.
•
Before putting into operation for the first time, check whether the
direction of rotation of the motor (fan direction of rotation) complies
with the direction arrow on the belt cover.
Содержание ABP 250
Страница 2: ...2 Copyright Copyright by Retsch GmbH Retsch Allee 1 5 42781 Haan Germany...
Страница 5: ...Notes on the Manual 5...
Страница 16: ...Technical Data 16 3 13 Installation drawing...
Страница 55: ...55...
Страница 56: ...56...
Страница 57: ...Copyright Copyright by Retsch GmbH Retsch Allee 1 5 42781 Haan Germany...