![Resvent RXiBreeze Скачать руководство пользователя страница 34](http://html1.mh-extra.com/html/resvent/rxibreeze/rxibreeze_user-manual_1445548034.webp)
RXiBreeze PAP System User Manual
32
15 Specification
Physical
Dimension (L*W*H): 238*178*128 mm
Weight: Approximately 1.60 kg
Operating Environmental
Temperature: 5
℃
~35
℃
Relative Humidity: 10%~95% (non-condensing)
Atmospheric Pressure: 70 kPa~106 kPa
Storage Environmental
Temperature: -25
℃
~70
℃
Relative Humidity: 5%~95% (non-condensing)
Atmospheric Pressure: 70 kPa~106 kPa
Noise Value
Sound pressure level Uncertainty Sound power level Uncertainty
30 dB(A)
2 dB(A)
38 dB(A)
2 dB(A)
Standards Compliance
IEC60601-1
:
2005+A1
:
2012 Medical electrical equipment - Part 1: General
requirements for basic safety and essential performance
IEC60601-1-2
:
2014 Medical electrical equipment - Part 1-2: General
requirements for basic safety and essential performance - Collateral Standard:
Electromagnetic disturbances - Requirements and tests
ISO 80601-2-70
:
2020 Medical electrical equipment
—
Part 2-70: Particular
requirements for the basic safety and essential performance of sleep apnoea
breathing therapy equipment
ISO 80601-2-74:2021 Medical electrical equipment
—
Part 2-74: Particular
requirements for basic safety and essential performance of respiratory
humidifying equipment
IEC60601-1-11
:
2015 Medical electrical equipment - Part 1-11: General
requirements for basic safety and essential performance - Collateral Standard:
Requirements for medical electrical equipment and medical electrical systems
used in the home healthcare environment
Electrical
Specified power supply: Resvent Medical Technology Co., Ltd.
Model: MDA60-220S24-W1
AC Input: 100-240V
~
1.18-0.56A 50/60Hz 2A Max
DC Output: 24 V
⎓
2.5 A
Specified power supply: Shenzhen Longxc Power Supply Co.,LTD
Model: LXCP61(
Ⅱ
)-024300
AC Input: 100-240
~
50/60Hz 1.5A Max
DC Output: 24.0V
⎓
3.0A