Restek Click-On Inline Super Clean Purifier Скачать руководство пользователя страница 2

1. Installation of a 

Click-On Connector

A Click-On connector is available for gas line tubing with OD of 

1

/

4

" and 

1

/

8

". These instructions are valid for each dimension of tubing; Restek 

advises the use of stainless steel or copper tubing for gas lines, preferably.

1.   Shut off the incoming gas.
2.  If no trap was previously in use, cut the gas line using an appropriate 

metal tubing cutter.

3.  Disconnect the gas line from the analytical instrument.
4.  Turn the carrier gas on and flush the gas line for 2 min with pure car-

rier gas; advised pressure: 90 psi.

5.  Remove the blue closing plugs at both ends of the trap and connect 

the Click-On connectors tightly to both ends of the gas line.

6.  For stainless steel tubing, always use stainless steel Click-On  

connectors; for copper tubing, always use brass Click-On connectors.

7.  Flush the incoming gas line and connectors for 10 min with pure  

carrier gas by pushing back the needle valve on the Click-On connector. 

NOTE: SHARP EDGE!! Advised pressure: 90 psi.

8.  Attach Click-On Trap to both Click-On connectors by punching the 

internal membranes in the traps and tightening the knurled ring 

clockwise.

9.  Install outlet connector first; then install inlet connector.
10. Flush the trap/whole system for at least 15 min with pure carrier gas; 

advised pressure: 90 psi (except for He-specific traps; they are already 

conditioned and need no flush). Do NOT connect the gas line to the 

analytical instrument yet!

11. Once the purifier has been flushed, mount the outgoing carrier gas 

line to the analytical system and you are ready to go: normally, the 

user should not see any notable argon or nitrogen peaks in the 

chromatogram.

12. Check whole system for leaks after installation!

2. Replacement of a Saturated 

Click-On Purifier

1.  A purifier cartridge needs to be exchanged when the visual indicators 

(if applicable) start changing color. The original indicator color and 

the color after saturation are printed on the front of the purifier.

2.  For metal traps without an indicator, it is strongly recommended you 

replace the trap once a year.

3.  Before installation of a new replacement trap, make sure that the 

temperature of the replacement trap is more or less identical to the 

temperature of the (already installed) Click-On connectors on your 

gas line to avoid any leakage in connections. If the temperature is not 

identical, use at least one hour of stabilization time in the same room 

as the connectors.

NOTE: 

the data given in the table strongly depend on the level of gas 

purity prior to entering the purifier.
Replacement of a metal purifier cartridge without an indicator is  

recommended once a year.

Brass Click-On 

Connectors

Stainless Steel 

Click-On Connectors

Click-On Inline Super Clean® purifiers are suited for the purification of 

noncorrosive gases with low contamination concentrations to a better 

as 6.0 grade (99.99990%) purity. Click-On Inline Super Clean® purifiers 

are connected to an appropriate genuine Click-On connector, which can 

only be connected to the gas line once.

CAUTION

  The maximum concentration of oxygen allowed in the gas before   

  entering the indicating trap is 0.5%. (Valid only for each purifier    

  packed with O

2

-adsorbents and O

2

-indicator.)

  Maximum gas-maximum system pressure must not exceed  

  11 bar (160 psi).

  Install outlet connector first; then install inlet connector.

  Do not install a purifier system near or in a hot area 

  (i.e., the hot-air exhaust at the rear of the GC).

  Never recondition any trap, ever: for eventual recycling, please    

  contact your local supplier.

  Always install traps in the appropriate position with the label at    

  the front.

  Installation of the connector to the gas line should be done by    

  experienced and/or certified personnel.

  Do not  forget to perform a reliable leak test, especially  

 

  when installing an indicating trap.

  During installation of a new Click-On Inline Super Clean®  

 indicating 

purifier, torsion on the plastic outer protection tube 

  should be avoided at any time; watch out for cracks in the inside    

  glass tube after installation (valid only for glass Click-On types  

  of traps).

  The screws on the outside trap housing on both sides should  

 

  never be removed; in case these screws should be missing, the 

  cartridge should not be installed; it should be returned to  

  the supplier.

  Click-On connectors are available as spare parts 

  – do NOT repair damaged connectors; replace them

  with new ones.

  Only use original parts to avoid malperformance and loss  

  of warranty.

NEVER INSTALL AN OXYGEN PURIFIER IN A GAS LINE  FOR AIR: 

it will result in instant saturation!

NOTE:

•  Once the purifier is positioned to the Click-On connector, the  

PTFE seals inside the purifier will be punctured, and it is necessary 

to keep the trap in position by hand to prevent air entering  

the system.

•  After installing the trap to the Click-On connector, highly pure gas 

will automatically stream from the purifier into the analytical system.

•  Regularly check both the big and small Viton O-rings on the 

Click-On connector for hair cracks. The slightest leak may allow 

moisture, oxygen, and hydrocarbons to enter the system and 

contaminate the system. Also, check the glass indicating purifier 

cartridge for cracks with a leak detector.

•  Indicator color change with moisture adsorbents devices 

according to EC Directive December 1998.

•  All purifiers are packed under argon, except for the special, 

He-specific GC-MS filter.

•  As long as there is no trap connected, the special Click-On 

connector will automatically block the gas stream.

NO GAS WILL PASS THROUGH THE GAS LINE!

Specific Capacities of 
Click-On Inline Super Clean® Purifier Types 

Purifier Type

Gas 

Quality at 

Outlet

Max.  

Pressure

Max.

Flow

Usable for:

Moisture

> 6.0

11 bar - 

160 psi

25 L/min

Inert carrier 

gas: He

*Air   *H

2

Oxygen

> 6.0

11 bar - 

160 psi

25 L/min

Inert carrier gas 

Hydrocarbons

> 6.0

11 bar - 

160 psi

25 L/min

Inert carrier 

gas: He

*Air   *H

2

Ind. Triple  

(mo 

  

hydrocarbons)

> 6.0

11 bar -  

160 psi

25 L/min

Inert carrier 

gas: He

Combi. (mo 

hydrocarbons)

> 6.0

11 bar - 

160 psi

25 L/min

Inert carrier gas: 

He *Air   *H

2

Triple (mo 

  

hydrocarbons)

> 6.0

11 bar - 

160 psi

25 L/min

Inert 

carrier gas 

Combi. (oxygen/

moisture)

> 6.0

11 bar - 

160 psi

25 L/min

Inert carrier gas: 

He, H

2

Ind. combi.  

(oxygen/moisture)

> 6.0

11 bar - 

160 psi

25 L/min

Inert carrier gas: 

He, H

2

Purifier Type

H

2

O (gr)

Capacity 

O

2

 (ml)

Hydrocarbons 

(g)

Estimated 

Lifetime

Moisture

21

n/a

n/a

> 3 years

Oxygen

n/a

3,000

n/a

> 3 years

Hydrocarbons

n/a

n/a

36  

(as 

n

-butane)

> 3 years

Ind. Triple  

(mo

  

hydrocarbons)

3

400

5  

(as 

n

-butane)

> 1 year

Combi. (mo 

hydrocarbons)

10

n/a

18  

(as 

n

-butane)

> 2 years

Triple (mo 

  

hydrocarbons)

6

1,000

12  

(as 

n

-butane)

> 2 year

Combi. (oxygen/

moisture)

10

1,500

n/a

> 2 year

Ind. combi.  

(oxygen/moisture)

3

500

n/a

> 2 year