Chapter 5
5-28
Esprit® Ventilator Service Manual © Respironics, Inc.
REF 580-1000-02 G
Repair
2. Connect the analyzer’s positive low pressure port to the patient circuit
Figure 5-17: Inhalation and Exhalation Pressure Transducer Troubleshooting Setup
3. Set the analyzer to read cmH
2
O.
4. Touch
Safety
to energize (white background) the safety solenoid.
5. Touch
Exhalation
and set it to 1470 steps.
6. Touch
Air
and set the flow to 1 LPM.
7. Touch
Exhalation
and adjust the steps until the analyzer pressure
reads 100 ± 5 cmH
2
O.
8. Check that the
Inhalation Pressure
and
Exhalation Pressure
displays
on the hardware screen read within ± 10% of the analyzer display.
9. Unblock the tee and verify that the inhalation, exhalation, and
analyzer pressure displays read 0 ± 0.1 cmH
2
O.