background image

Lab

RAM II Jacketed Vessel Users’ Manual  

 

REPLACEMENT PARTS 

 

 

 

Resodyn Acoustic Mixers, Inc. 

010320 

 

General Notes 

For use with the LabRAM ll mixer, the Jacketed Vessel assembly consists of: 

  Jacket fixture that attaches to the LabRAM II Resonator Plate  

  Removable mixing vessel 

  Fixture lid with RTD and vacuum ports.  

The system bolts into place, replacing the vessel holder on the LabRAM ll and features 

ports for an RTD temperature probe port and two vacuum lines. The system’s metallic 

surfaces are 316 stainless steel.  

4.1 

System Description 

The LabRAM II Jacketed Vessel accessory consists of the following: 

  One 316 stainless steel (SS) Jacketed Vessel fixture with 1/8” female NPT 

connections for heating/cooling transfer fluid inlet and outlet 

  One 500 ml/16 oz. payload capacity 316 stainless steel mixing vessel  

  One 316 stainless steel fixture lid with fittings for vacuum lines and RTD probe  

  One stainless steel Sanitary Quick-Clamp 

  Four Viton O-rings (two large and two small), required for face seal between 

Jacketed Vessel fixture and removable mixing vessel, and between the removable 

mixing vessel and Jacket fixture lid.  
For O-rings of special materials, please call 406-497-5333 

  6” long 316 SS sheathed Resistance Temperature Detector (RTD) 

  PTFE RTD sleeve, ½” PTFE front and rear ferrules, and compression nut 

  RTD Cable  

  Vacuum lines connecting LabRAM II vacuum ports to Jacketed Vessel fixture Lid 

  Four tube plugs (for operation without vacuum, includes two extra) 

  Four bolts with washers for anchoring the Jacketed Vessel fixture to the LabRAM II 

platform 

  Two braided SS flex hoses for Jacketed Vessel heating/cooling transfer fluid 

service loop connections to back panel of LabRAM II  
 

A cutaway of the LabRAM II Jacketed Vessel system including significant components is 

shown in Figure 4-1.  

 

Note: A compatible LabRAM II mixer and heat/cooling transfer fluid supply system are not 

included in the Jacketed Vessel components, but are required for its use. Refer to the 

user manual for your heater/chiller system for installation and operating instructions. If you 

need a heat transfer supply system, please contact Resodyn at 406-497-5333. 

 

Содержание LabRAM II

Страница 1: ...Resodyn Acoustic Mixers Inc 130 N Main Street Suite 630 Butte MT 59701 406 497 5333 www resodynmixers com LabRAM II Jacketed Vessel Users Manual ...

Страница 2: ...tant Safety Notes 5 4 General Notes 7 4 1 System Description 7 5 Installation 9 5 1 Jacketed Vessel Assembly Component Pictures 9 5 2 Jacketed Vessel Accessory Installation and Removal 11 6 Operating Instructions 13 6 1 LabRAM II Jacketed Vessel Accessory Operation 13 7 Service Information 14 7 1 Maintenance and Service 14 8 Supplemental Information 15 8 1 Replacement Parts 15 ...

Страница 3: ... additional accessories or replacement parts Key features of the LabRAM II Jacketed Vessel Control of fluid temperature from 0 F cooling to 302 F heating 18 C to 150 C Facilitates mix quality when temperature control is required Access ports to enable use of the vacuum processing option Enables temperature monitoring with access port for RTD hardware Removable replaceable mixing vessel liner in mu...

Страница 4: ...Mixing Power The amount of power being delivered by the RAM system to perform mixing Mixing Regime Specific modes of mixing For example a fluid will change regimes as it undergoes changes in acceleration from an incompressible liquid containing few headspace bubbles to a fluid that is fully integrated with the compressible headspace such that the two are indistinguishable from each other PTFE Poly...

Страница 5: ...0 psig 1 to 4 5 bar absolute jacket pressure Suitable for intermittent or continuous operation For use in non hazardous environments or if used with the LabRAM II H mixing platform in hazardous environments as classified by the United States National Electrical Code NEC Class I Div I and Class II Div I 3 2 Warnings and Cautions Throughout the manual the following symbol is used to identify warning...

Страница 6: ...nstalling tubing or opening the Jacketed Vessel CAUTIONS Use recommended fittings only Use of improper fittings and or improper connection to fittings may result in leakage at the fittings A user supplied heat transfer fluid circuit pressure relief bypass valve must be installed in a potentially high pressure system to prevent the pump from exceeding the maximum rated pressure i e 50 psig 3 5 bar ...

Страница 7: ... two small required for face seal between Jacketed Vessel fixture and removable mixing vessel and between the removable mixing vessel and Jacket fixture lid For O rings of special materials please call 406 497 5333 6 long 316 SS sheathed Resistance Temperature Detector RTD PTFE RTD sleeve PTFE front and rear ferrules and compression nut RTD Cable Vacuum lines connecting LabRAM II vacuum ports to J...

Страница 8: ...10119 Figure 4 1 Jacketed vessel cutaway showing the components included in the assembly except hoses and cables RTD Assembly Jacketed Vessel Lid Vacuum Port 1 of 2 Sanitary Quick Clamp Heat Transfer Fluid Port Bolt for Mounting on LabRAM II O ring Face Seals Mixing Liner Jacketed Vessel Jacket ...

Страница 9: ...ble mixing vessel in which mix materials are added and mixed and the Jacketed Vessel fixture lid with integrated RTD and vacuum ports In Figure 5 2 the fully assembled Jacketed Vessel is pictured mounted on a LabRAM II mixer resonator plate with all connections attached Note that each line or cable attached to the Jacketed Vessel fixture forms a gradual arc or service loop to accommodate the verti...

Страница 10: ... Vessel Users Manual SAFETY INFORMATION Resodyn Acoustic Mixers Inc 10 Rev 110119 Figure 5 2 Jacketed Vessel mounted on a Resodyn LabRAM II mixer Figure 5 3 RTD assembly sleeve nut and ferrules used to seal the RTD feed ...

Страница 11: ...Attach each of the Heat Transfer Fluid Hoses to the back panel of the LabRAM II mixer as shown in Figure 5 2 The Jacketed Vessel should work equally well regardless of the direction of flow but if a specific flow direction is desired note the hose locations Place the removable Mixing Vessel on top of the now mounted and installed Jacketed Vessel fixture Place an O Ring in each of the two O Ring gr...

Страница 12: ...up or down relative to the Jacketed Vessel lid during resonant acoustic mixing at high i e 100 g accelerations If process vacuum is to be used loosen the white compression nuts in the back panel and thread each of the Vacuum Lines from the vacuum unit through the back of the panel to form a gradual 90 degree arc service loop down to the Jacketed Vessel fixture lid as shown in Figure 5 2 Attach the...

Страница 13: ...ssure if used and heating cooling transfer fluid supply pressure e g turn off the heat transfer fluid pump If the Jacketed Vessel is not isolated from the heating cooling transfer fluid supply pressure before unclamping the Jacketed Vessel transfer fluid will leak out onto the mixer platform Loosen and remove the Sanitary Quick Clamp and remove the Jacketed Vessel fixture lid from the Jacketed Ves...

Страница 14: ... remove the Jacketed Vessel fixture lid from the removable Mixing Vessel to recover the processed material Take care not to contaminate the materials with the transfer fluid from the underside of the Mixing Vessel liner 7 Service Information 7 1 Maintenance and Service The LabRAM II Jacketed Vessel accessory should only be operated when it is in good working condition If the system shows any signs...

Страница 15: ... O Ring Small Viton 6 4 A01A 0250 20 00750 ABB 1 4 Hex Head Bolt 7 1 102760 Mixing Liner 16 oz Stainless 8 1 102761 Jacketed Vessel Lid 9 1 102784 RTD PTFE Sleeve 10 1 3305 O Ring Small Viton 11 1 3306 O Ring Large Viton 12 1 903345 RTD Cable not shown 13 1 903324 RTD Assembly 6 Probe 14 1 2956 1 2 PTFE Ferrule Set 15 4 102665 Tube Plug PTFE 16 2 1339 Heat Transfer Fluid Hose for Service Loop not ...

Страница 16: ...Innovative Mixing Technologies 130 North Main St Suite 630 Butte MT 59701 406 497 5333 info resodynmixers com www resodynmixers com LabRAM LabRAM II LabRAM II H OmniRAM RAM 55 RAM 5 ...

Отзывы: