ResMed VPAP IV ST Скачать руководство пользователя страница 5

ACDuncan Dec 2008 
Reviewed 2013

 

 

 

VPAP TROUBLESHOOTING GUIDE 

Problem         

 

 

Possible Cause 

 

 

Corrective Action______ 

Unit does not start when the   

Power cord not connected properly  Connect cord firmly at 

Start button is pressed, and   

 

 

 

 

 

both ends. 

no lights are showing  on 

 

  

the LCD panel. 

 

 

Power point may be faulty 

 

Try another power outlet 

Machine does not alternate    

Hyperventilation and/or high   

The machine will revert  

between IPAP and EPAP 

 

room temperature has caused 

to normal once it cools   

 

 

 

 

 

machine to overheat   

 

down (about 15 mins) 

If  uncomfortable  stop  the 

machine  for  the  time  it 

takes to cool. 

 

 

 

 

 

 

 

 

 

 

Try to slow your breathing. 

 

 

 

 

 

 

 

 

 

 

Cool your room down. 

Machine not giving enough   

Ramp timer is on 

 

 

Wait for pressure to build  

air 

 

 

 

 

 

 

 

 

 

up over selected ramp time  

 

 

 

 

 

 

 

 

 

 

or Turn ramp timer off. 

 

 

 

 

 

Air filter is dirty 

 

 

Change the filter. 

 

 

 

 

 

Tubing is kinked or punctured 

Straighten or replace tube. 

 

 

 

 

 

Tubing is not connected properly 

Tighten all connections  

 

 

 

 

 

Mask and headgear not on properly  Reapply mask 

 

 

 

 

 

Plugs missing from access ports 

Replace plugs. 

 

 

 

 

 

on mask. 

Air leaking around nose 

 

Cushion not properly fitted to  

Refit cushion to frame. 

and eyes 

 

 

 

mask frame 

 

 

 

 

 

Mask/harness not positioned  

Adjust mask position. 

 

 

 

 

 

correctly. 

 

 

 

 

 

Plugs missing from access ports 

Replace plugs or seal hole 

 

 

 

 

 

on the mask   

 

 

with plastic tape. 

 

 

 

 

 

Nasal cushion is worn, torn or 

Replace cushion 

 

 

 

 

 

distorted. 

Machine does not stop when you  

Smart start/stop is disabled   

Enable Smart Start/Stop. 

take the mask off 

 

 

 

 

 

 

 

Use on/Off button to stop 

 

 

 

 

 

 

 

 

 

 

the machine 

Smart Start/stop is on but unit 

Incompatible humidifier or 

 

Change humidifier or mask 

does not stop when you take the 

mask being used 

 

 

type 

mask off 

Displays error message:

 

 

The air tubing is loose 

   

Secure all connections 

Check tube!!   

 

 

 

 

 

 

 

Press any key to clear the 

Key if done. 

 

 

 

 

 

 

 

 

message. 

Displays error message:

 

 

Component failure 

 

 

Call VRSS Outreach 

SYSTEM ERROR 

Call Service! 

Excessive motor noise 

 

Component failure 

 

 

Call VRSS Outreach   

Displays error message:

 

           You had excessive mask/ 

 

Secure all connections 

High leak in last session 

 

mouth leak during the night   

Use mask fitting feature to  

 

 

 

 

 

 

 

 

 

 

help fit mask properly. 