HumidAire 2i

 

HEATED HUMIDIFIER

Disinfection Guide

English

 

Содержание HumidAire 2i

Страница 1: ...Clinical Use Only 1 HumidAire 2i HEATED HUMIDIFIER Disinfection Guide English ...

Страница 2: ...acility consult its own procedures before carrying out the following instructions WARNINGS and CAUTIONS ResMed cannot give any assurance that deviations from the procedures listed in this guide and their effect on the performance of the product will be acceptable When using detergents or disinfectants always follow the manufacturer s instructions Note The HumidAire 2i is available in both a clear ...

Страница 3: ...er approximately 113 F or 45 C according to the manufacturer s instructions Drying Not applicable Pre disinfection Not applicable Drying Not applicable Disinfection high level thermal disinfection 1 Activate the automatic washer disinfector for a standard cycle of thermal disinfection2 2 Ensure that the machine has thoroughly rinsed each component 2 Disinfection Due to specific regional requiremen...

Страница 4: ...level chemical disinfection 1 Soak the disassembled water chamber components in a commercially available solution of ortho phthalaldehyde 0 55 eg CIDEX OPA 2 Take care that no air bubbles are trapped against the components 2 Thoroughly rinse3 each component according to the manufacturer s instructions 2 Disinfection ResMed water chambers have been validated for 20 cycles using solutions of ortho p...

Страница 5: ...hamber To do this lift the latch at the bottom of the water chamber and pull the water chamber away from the docking station 4 Discard any excess water by turning the water chamber upside down allowing the water to drain out of the air outlet 5 Disassembling the Water Chamber 1 To open the side catches pull out from the lower edge 2 Separate the water chamber by pulling the two halves apart Use yo...

Страница 6: ...the base 4 Disengage the catches from the sides of the water chamber base Then separate the inner and outer moulding of the catch pull back the inner moulding and squeeze The inner moulding will pop out 7 Reassembling the Water Chamber When all the water chamber parts are clean and dry reassemble the water chamber 1 Reassemble the catches by inserting one side of the inner moulding to the outer mo...

Страница 7: ...bs Push firmly and click into place 4 Refit the O ring in the channel of the tub plate CAUTION Ensure that you replace the O ring if it is nicked or punctured 5 Push the tub plate into the water chamber with the open side facing down Then firmly press the tub plate down into the water chamber 6 Place the cross member on its supports inside the water chamber base Tub plate ...

Страница 8: ...r 1 Elizabeth Macarthur Drive Bella Vista NSW 2153 Australia ResMed Corp US Designated Agent 14040 Danielson Street Poway CA 92064 6857 USA ResMed UK Ltd EU Authorized Representative 96 Milton Park Abingdon Oxfordshire OX14 4RY UK ResMed Offices Australia Austria Brazil China Finland France Germany Hong Kong India Japan Malaysia Netherlands New Zealand Norway Singapore Spain Sweden Switzerland UK ...