background image

 

V4: August 2018 

 

Page 26 of 31 

For Review: August 2019 
Author: LTV Team/JT 

Understanding Alarms 

 

Low VTE/Low MVE

 

Disconnection/leak

 

 

1 Assess child 

Accidental 
decannulation 

(i.e. tracheostomy 
has come out) 

Action 

Immediately insert 
tracheostomy 
If difficulty follow 

emergency 
algorithm

 

1 Assess child 

Possible 
causes 
include: 

Blocked 
tracheostomy

 

Action 
Emergency 
algorithm: 

1. Suction 
2. Emergency 
tracheostomy 
change 

2 Assess 
equipment 

Disconnection 
within circuit 

(follow circuit from 
child through to 
ventilator and 
ensure everything 
is connected 

 

NB humidifier 
connectors etc 
may be slightly 
loose) 

Cause 

- some leak may be 
tolerated and may 
be due to position 
of child 

 discuss 

with community ± 
medical teams, who 
can consider need 
for cuffed trache or 
upsize 

2 Assess child 

Retained 
secretions/ 
increased 
pulmonary 
resistance etc

 

Action 

• suction,

 

consider need 
for nebulisers, 
physiotherapy 

• refer to advanced

 

treatment plan 

• discuss with

 

community or 
medical team 

3 Assess child 

Possible 
causes include: 

Leak around 
tracheostomy 

(particularly when 
asleep) 

Action 

• Reconnect any

 

loose connections 

• Re

-assess 

3 Assess 
equipment 

Circuit 
blockage 

(Follow circuit 
from child through 
to ventilator 
and ensure it is 
not kinked or 
obstructed) 

1 Assess child 

• Correct fault

 

• Re

-assess 

Is the alarm set 
appropriately 

(i.e. as previously 
recorded and 
checked at start of 
shift) 

• Check alarm

 

settings are as 
prescribed and 
re-set if any 
discrepancies 

Is the alarm set 
appropriately 

(i.e. as previously 
recorded and 
checked at start of 
shift) 

• Check alarm

 

settings are as 
prescribed and 
re-set if any 
discrepancies 

If the child is compromised and you can’t resolve the issue rapidly

 

 hand ventilate and call for help

 

 

 

 

 

 

 

 

 

 

Содержание Astral 150

Страница 1: ...ildren s Long Term Ventilation Service Clinical Team Email HospitaltoHomesupport rbht nhs uk Telephone 0207 352 8121 x 84411 Astral 150 Ventilator Information and User Guide for Qualified Professional...

Страница 2: ...age 23 Setting Alarms Page 23 Activated Alarms Page 24 Testing Alarms Page 25 Understanding Alarms Page 26 Troubleshooting Common Problems Page 27 Using Nebulisers Page 29 Consumables Page 31 Ventilat...

Страница 3: ...Author LTV Team JT Description of Outside of Ventilator Inspiratory Port Air going to the patient Expiratory Port Expired air coming back to the ventilator Power Button USB Connector to download moni...

Страница 4: ...V4 August 2018 Page 4 of 31 For Review August 2019 Author LTV Team JT 1 Touch screen 2 Power source indicator 3 Ventilation on off indicator 4 Alarm mute button 5 Alarm light...

Страница 5: ...menu 2 Manual breath button only shown if enabled and in some settings 3 Information bar see page 6 for more information 4 Battery power indicator 5 Lock touch screen button 6 Menu bar see page 6 for...

Страница 6: ...is displayed at the top of the touch screen The Information bar displays the operating status of the device including patient type current circuit configuration programs information messages ventilati...

Страница 7: ...y power external battery power and car charger power source Removing the power lead The power cord is equipped with a push pull locking connector To remove grasp the power cord housing and gently pull...

Страница 8: ...tor parameters and the child s respiratory pattern Please observe the internal battery indicator carefully Warning Please ensure to regularly check the internal battery The internal battery should be...

Страница 9: ...y pattern Please observe the internal battery indicator carefully The external battery cannot be charged from your main power This can be used and recharged while connected to the Astral device Previo...

Страница 10: ...Team JT Using a car Charger When using a car adapter start the car before connecting the adapter If the power source drops the ventilator will switch automatically to internal battery power Types of...

Страница 11: ...e i e once every 7 days See local hospital guideline for antibacterial filter Order code 7073800 MR7LEAK15 from ResMed for circuit 24988 from Resmed for additional leak valves 15 mm circuit See equipm...

Страница 12: ...from Intersurgical for HME See equipment list for other order code Other Circuit that may be seen Double Limb Circuit wet The whole circuit should be changed weekly at home including antibacterial fil...

Страница 13: ...ode 5500 from Intersurgical Note Heat Moisture Exchanger should be changed daily i e once every 24hours Order code 1644 from Intersurgical for filter Turning On Off To turn the machine on and off pres...

Страница 14: ...ach programme and the ventilator will then remember the test for the relevant circuit when the programme is selected From the Setup Menu select the Circuit sub menu Press start and follow the on scree...

Страница 15: ...nutes or Unlimited then it will direct you to the main settings You will be able to change the settings on this interface depending on the tab you are in To exit clinical menu press The exit clinical...

Страница 16: ...ll display then select Continue Select the type of patient adult or paediatric Select the circuit type you wish to use The ventilator will now prompt you to run a learn circuit see page 14 Following a...

Страница 17: ...s highlight each parameter in turn As a parameter is highlighted up and down arrows will appear to the right hand side of the screen Use these to select the desired value Once you have finished press...

Страница 18: ...lation Programmes Up to 4 ventilation programmes may be enabled To do this unlock the clinical menu press and hold the padlock symbol for 3 seconds Select 20 minutes or Unlimited then it will direct y...

Страница 19: ...Ventilation Programmes It is possible to set up more than one ventilation programme For example some children may have a programme of ventilation set to use when they are well and another for when the...

Страница 20: ...icon To stop the ventilation press and hold the stop ventilation icon for 3 seconds Release when prompted and then select confirm Entraining Oxygen Oxygen is entrained via the oxygen nozzle on the ba...

Страница 21: ...ange the filter even if the filter is less than 4 weeks old NB The filter is not reusable and cannot be washed Before replacing the air filter turn off the device and remove mains power and or externa...

Страница 22: ...low or the device is powered down Caution Only select unlimited if the ventilator will be under constant observation of a suitably qualified professional To lock press the padlock icon then press conf...

Страница 23: ...ight hand side of the screen Select Settings then highlight the parameter you wish to alter As a parameter is highlighted up and down arrows will appear to the right hand side of the screen Use these...

Страница 24: ...priority alarm or the last active alarm not yet reset 2 Active Alarms Screen Displays all the active alarms Alarms will disappear from this list as they are resolved 3 Information Menu To see a histo...

Страница 25: ...ing Alarms Alarm Checks Must be carried out at the beginning of each shift A good time to do this would be when the patient requires suctioning It would be useful for two people to do this in order to...

Страница 26: ...or upsize 2 Assess child Retained secretions increased pulmonary resistance etc Action suction consider need for nebulisers physiotherapy refer to advanced treatment plan discuss with community or me...

Страница 27: ...being held with the front of the ventilator facing the floor Humidification Problems Humidifier will alarm if temperature is too high or too low a Heater wire life spam approximately 8 months b Temper...

Страница 28: ...cuit Check circuit for splits check for disconnection at the tracheostomy of the humidifier wires and at the machine Ventilator Alarming High Low Respiratory Rate May be due to a Respiratory limits se...

Страница 29: ...below The nebuliser is powered by an air compressor Using a Nebuliser in a Double Limb Circuit The Double limb circuit is a closed system therefore particles will remain within the system Using a Nebu...

Страница 30: ...s in a single leaked circuit an antibacterial filter must be used in between the nebuliser chamber and expiratory leak to prevent particles escaping into the atmosphere Please note the correct filter...

Страница 31: ...alarms prior to connecting the child to the ventilator Good practice to perform the Learn Circuit within the last month Ensure that you are in the correct programme as per individualize patient s pla...

Отзывы: