XCell™ C410:V3 Controller
User Guide
14 repligen.com UG-C410v3-06
2.
XCell™ C410:V3 Controller layout
2.1
General layout
The XCell™ Controller consists of two parts: The P-Box,
Figure 4
Figure 5
. The two are
interconnected with a cable that relays signal and power. A general layout of the two boxes with the
XCell ATF® Device is shown in
. The primary design objective is to produce a modular system
that will maximize adaptability of the system to the various space requirements of the user’s
facilities. The Interconnect Cable, I-Cable, connecting the E-Box and the P-Box can be selected to the
required lengths to facilitate positioning and handling of the boxes within the facility. One can
envision the P-Box in proximity to the filtration assembly, while the E-Box positioned distant to the
filtration assembly, possibly mounted on a wall or a skid.
2.1.1
Filtration assembly
This includes the diaphragm pump , Filter Housing, connection to the bioreactor, harvest line, pump
air inlet assembly, stand, plus all the housing accessories as specified in the part list.
Figure 4. Pneumatic Enclosure
a.
System STOP switch
b.
Interconnect cable plug
c.
Plugs for sensor inputs (4x4-20mA)
d.
Connection for diaphragm pump
e.
Connection for compressed air
f.
Connection for vacuum supply
g.
Pressure regulator (0 to 60 psi, 0 to 4.1 bar)
h.
Vacuum gauge (0 to -14psi, o to -0.95 bar)
i.
Pressure gauge (0 to 60psi, 0 to 4.1 bar
(Not pictured in
) Pressure relief assembly - set to 40 psi, 2.8 bar.
Electric enclosure
The Electric enclosure contains the HMI and PLC components, including: