TONiC RTLC20-S Functional Safety installation guide and safety manual                      Original instructions

RTLC20-S scale installation drawing

Dimensions and tolerances in mm

Scale length ( L )

Overall length ( L + 37 )

Measuring length ML = (L − 17)  (ML = (L − 32) with dual limits)

Readhead optical detector position at extent of travel

P limit magnet
(A-9653-0138)

P and Q limit switch 

sensor position

TONiC readhead

Nominal P limit 

trigger point

Q limit magnet

(A-9653-0139)

(Dimensions as P limit)

Reference mark selector magnet

(A-9653-0143)
(Dimensions as P limit)

8

B

Optical centreline (incremental and reference mark)

5

10

9.2

15

RTLC20-S scale

Datum clamp 

(A-9585-0028)

End cover
(Pair A-9585-0035) 

5

STAR

(see page 8)

FINISH 

(see page 8)

A

C

Detail B

End cover 

alignment 
indicator

Detail C

0.5

F

0.2/100

0.05

F

F = axis of motion

F = axis of motion

Nominal Q limit 

trigger point

Ra 3.2

Detail A 

Optional bolted reference mark selector and limit magnets

22

18

10

3.7

Ø2.2

 

*

 Dimension from substrate surface.   

 Supplied with 2 × M2 × 4 screws.

1.85

3.7

1.5 *

4.5

12

15 ±1

18.5 ±

0.8

3.5

(adhesive datum clamp)

NOTES:

  The reference mark selector and limit actuator locations are correct for the readhead orientation shown.

 

External magnetic fields greater than 6 mT, in the vicinity of the readhead, may cause false activation of the limit and reference sensors. 

Bolted magnet type

Part number

P limit

A-9653-0292

Q limit

A-9653-0291

Reference mark selector

A-9653-0290

Содержание TONiC FS T3 Series

Страница 1: ...Installation guide M 6688 9044 02 B TONiC FS T303x RTLC20 S linear encoder system Functional Safety installation guide and safety manual...

Страница 2: ...TONiC FS quick start guide 10 Cable connection 11 System connection Ti interface 11 System connection DOP interface 12 Readhead mounting and alignment 13 System calibration 14 Diagnostic LEDs 15 Outpu...

Страница 3: ...orrect the interference at their own expense NOTE This unit was tested with shielded cables on the peripheral devices Shielded cables must be used with the unit to ensure compliance Further informatio...

Страница 4: ...rer must conduct their own assessment of the full system to determine its safety capability Correct use includes X X Operating the TONiC FS system within the limits defined in this document X X Instal...

Страница 5: ...als falls within the limits defined in Evaluation unit monitoring Evaluation unit monitoring To achieve full system integrity the evaluation unit must continuously monitor the analogue outputs and in...

Страница 6: ...79 The products comply with EU directives and comply with the following technical standards 2014 30 EU Electromagnetic compatibility EMC 2011 65 EU On the restriction of the use of certain hazardous s...

Страница 7: ...s RTLC20 S 150 mm NOTE Ensure self adhesive tape is on the outside of bend 95 relative humidity non condensing to IEC 60068 2 78 Humidity Storage 70 C 0 C Operating Storage and handling N heptane CH3...

Страница 8: ...ce mark selector sensor position P and Q limit switch sensor position Selected IN TRAC reference mark 0 08 Roll tol 0 5 11 13 5 4 25 4 15 10 8 75 Offset 2 75 0 5 3 0 25 Yaw tol 0 4 29 22 Detail A Scal...

Страница 9: ...scale Datum clamp A 9585 0028 End cover Pair A 9585 0035 5 START see page 8 FINISH see page 8 A C Detail B End cover alignment indicator Detail C 0 5 F 0 2 100 0 05 F F axis of motion F axis of motio...

Страница 10: ...on There are two versions of applicator for use with RTLC20 S scale side mounted A 9589 0115 and top mounted A 9589 0094 NOTE Ensure the guidance in Storage and handling on page 5 is followed 9 Remove...

Страница 11: ...need to be fitted adjacent to a reference mark It can be positioned anywhere along the axis depending upon the customer s requirements 1 Remove the backing paper from the datum clamp Place the datum c...

Страница 12: ...by the readhead and interface set up LEDs readhead Green DOP interface ideally Blue Purple CALIBRATION Press and release the CAL button on the interface The CAL LED on the readhead will be single fla...

Страница 13: ...n the readhead and the interface X X The customer is responsible for verifying the product function where the readhead to interface cable has been re terminated including the installation of extension...

Страница 14: ...ing care not to touch the pins plug the connector into the socket in the interface ensuring correct orientation as shown 3 Refit the cover plate ensuring the cable ferrule is located in the recess on...

Страница 15: ...nce will contribute to mechanical safe position Green Orange Red Readhead set up LED status Yaw 0 0 4 Pitch 0 1 Roll 0 0 5 Rideheight 2 1 0 15 mm Green spacer DOP interface set up LED status Blue Gree...

Страница 16: ...stops flashing and remains off The reference mark is now phased NOTE Only the chosen reference mark that has been used in the calibration routine is guaranteed to remain phased X X The system automati...

Страница 17: ...l 50 to 70 No Red Poor set up signal may be too low for reliable operation signal level 50 No Purple blank flashing Over signal system in error Yes Blue blank flashing Over speed system in error Yes R...

Страница 18: ...Shield Inner shield Not connected Outer shield Case Function Output type Signal Colour Power 5 V Brown 0 V White Incremental signals Analogue Cosine V1 Red Blue Sine V2 Yellow Green Reference mark An...

Страница 19: ...2 70 1 35 0 540 0 270 0 135 0 054 0 027 25 10 10 8 10 3 24 1 62 0 810 0 324 0 162 0 081 0 032 0 016 20 10 10 6 75 2 70 1 35 0 675 0 270 0 135 0 068 0 027 0 013 12 10 9 4 50 1 80 0 900 0 450 0 180 0 09...

Страница 20: ...5 V Readhead IMPORTANT The outer shield must be connected to the machine earth Field Ground The inner shield must be connected to 0 V at receiving electronics only Care should be taken to ensure that...

Страница 21: ...lect standard or wide reference at time of ordering to match the requirements of the controller being used 20 m 45 Output specifications Analogue output signals Incremental 2 channels V1 and V2 differ...

Страница 22: ...ing 0 C to 70 C Sealing readhead IP40 Ti interface IP20 DOP interface IP30 Acceleration readhead Operating 500 m s 3 axes Shock system Non operating 500 m s 6 ms sine 3 axes Vibration system Operating...

Страница 23: ...octite 435 with A 9585 0028 clamp RTLC20 S scale technical specifications For lengths 2 m RTLC20 with FASTRACK is recommended Reference mark Limit switches Type Customer selected IN TRAC reference mar...

Страница 24: ...nable disable Push and release 3 seconds AGC enable disable Push and hold during power Off On cycle Restore factory defaults Refer to readhead LED functionality chart for CAL set up indications 48 32...

Страница 25: ...T 2019 Renishaw plc All rights reserved Renishaw reserves the right to change specifications without notice RENISHAW and the probe symbol used in the RENISHAW logo are registered trade marks of Renish...

Отзывы: