IMP Wires
Two separate insulated wires (each single core).
Outside diameter
1,0mm (0.039in).
Length
2000mm (78in).
Colours
Pink and White.
Warning - DO NOT pull or repeatedly bend the wires during installation.
dimensions mm (in)
IMP MOUNTING
IMP GENERAL ARRANGEMENT
IMP INSTALLATION
The IMP is secured using flat ended grubscrews
(over clamping zone marked x x ),
or a semi-permanent adhesive.
RECOMMENDED GRUBSCREW TORQUE
Clamping screw size
M4 maximum.
Clamping screw torque
0,25Nm (0.18 lbf.ft) maximum.
Warning - DO NOT use cone point screws for clamping.
RECOMMENDED ADHESIVE
Silicon RTV
DOW CORNING Q3-3744 or equivalent.
IMPORTANT
Pressure may be applied to the metal housing around
the circumference, but the plastic front face should not
be subjected to direct pressure on installation.
Hold with flat plate until adhesive cures.
20,1 (0.791)
20,0 (0.787)
15 (0.59)
MINIMUM
Clamping zone
15°
Ø21 (0.83)
0,8 (32)
Ø
4,0 (0.157)
2,0 (0.078)
Ø6,5
(Ø0.26)
MAX
Ø19,95
(Ø0.785)
MAX
13,25 (0.52)
13,05 (0.51)
REF
16,25 (0.64)
MAX