background image

IMP Wires
Two separate insulated wires (each single core).
Outside diameter

1,0mm (0.039in).

Length

2000mm (78in).

Colours

Pink and White.

Warning - DO NOT pull or repeatedly bend the wires during installation.

dimensions mm (in)

IMP MOUNTING

IMP GENERAL ARRANGEMENT

IMP INSTALLATION

The IMP is secured using flat ended grubscrews
(over clamping zone marked  x  x  ),
or a semi-permanent adhesive.

RECOMMENDED GRUBSCREW TORQUE
Clamping screw size

M4 maximum.

Clamping screw torque

0,25Nm (0.18 lbf.ft) maximum.

Warning - DO NOT use cone point screws for clamping.

RECOMMENDED ADHESIVE
Silicon RTV

DOW CORNING Q3-3744 or equivalent.

IMPORTANT
Pressure may be applied to the metal housing around
the circumference, but the plastic front face should not
be subjected to direct pressure on installation.

Hold with flat plate until adhesive cures.

20,1 (0.791)
20,0 (0.787)

15 (0.59)

MINIMUM

Clamping zone

15°

Ø21 (0.83)

0,8     (32)

Ø

4,0 (0.157)

2,0 (0.078)

Ø6,5

(Ø0.26)

MAX

Ø19,95

(Ø0.785)

MAX

13,25 (0.52)
13,05 (0.51)

REF

16,25 (0.64)

MAX