Mounting the probe to the probe head
If fitting a TP20 probe, mount the probe body to the probe head before fitting the probe module and stylus.
Screw the threaded end of the probe into the M8 connector on the probe head until it is finger tight
Fit the S1 ‘C' spanner ﴾supplied﴿ to the location holes in the probe body and tighten by hand
The recommended tightening torque is 0.3 Nm - 0.5 Nm
Refer also to the user's guide for the probe type being used.
Extension bars
Probe reach may be extended by 50 mm using a PEL1 extension bar. Use of the PEL2, PEL3 or PE series extension bars is not recommended.
MH8 user's guide
www.renishaw.com
Issued 02 2020
9