8.  Technical Data

Type

934 0100

935 0100

Gas type

Liquid gas

G 30 / 50 mbar

Natural gas E

G 20 / 20 mbar

Height

approx. 97 mm

[3.82 inch]

Weight

approx. 460 g

Connection hose

according to DIN 30664 or DVGW G5501-P

Connection pressure

50 mbar

[0.725 psi]

20 mbar

[0.290 psi]

Consumption

50 g/h

48 l/h

Watt

650 W

450 W

Nozzle Ø

0,34 mm 

[0.01 inch]

0,55 mm 

[0.02 inch]

Operating mode

Continuous operation

max. ambient temperature

40 °C [104 °F]

9.  To change the gas type 

The burner corresponds to the CE 30665 part 1 requirements and can be converted to use a different 

type of gas by changing the nozzle.

The type of gas and performance can be found in the technical data.

• 

The liquid gas version has 

“Liquid gas 50 mbar” 

written on the identification plate.

• 

The natural gas version has 

“Natural gas E 20 mbar”

 written on the identification plate.

When the nozzle is changed, the identification plate must also be changed.

The conversion sets can be found in the spare part list.

  The nozzle may only be changed by qualified personnel according to the DVGW or valid natio

-

nal regulations. 

10. Guarantee

All burners are checked before distribution. If you discover any functioning fault, please send the bur

-

ner in for repair with a description of the fault. Do not carry out any repairs yourself. The burner nozzle 

and mixing tube have been carefully aligned and are compatible with one another; they must not be 

altered as they would then no longer correspond with the regulations. 

When the equipment is used as intended and stipulated, Renfert warrants all parts of the unit with a 

guarantee of 3 years.

 

An original sales invoice from a specialist dealer is a precondition for a warranty claim.

Parts which are susceptible to regular wear and tear, such as consumables, are excluded from this 

warranty. 

The guarantee is invalid after improper use, incompliance with the instructions for use regarding ope-

ration, cleaning, maintenance and connection, after repairs which were not carried out by specialists, 

if spare parts from other manufacturers have been installed and unpredictable or unpermitted external 

influences. 

Guarantee claim do not lengthen the original guarantee duration. 

Содержание Safety Burner eco 934 0100

Страница 1: ...E Bedienungsanleitung Bedienungsanleitung Instruction manual Mode d emploi Istruzioni d uso Instrucciones para el servicio Instru es de uso Kullan m K lavuzu Datei 216715_01 Titel nderungsstand 20 08...

Страница 2: ...you will find these symbols with the following meanings Danger Immediate danger Observe the accompanying documentation Warning By not following the instructions the burner could become damaged Note Us...

Страница 3: ...g smell of gas be detected close the gas shut off valve at the laboratory work bench immediately and initiate the corresponding safety measures see DVGW law and valid national regulations Always pay a...

Страница 4: ...at the laboratory workbench gas isolation valve Slide the certified gas hose in Germany according to DIN 30664 or DVGW G 5501 P up to the last groove of the hose spout 5 on the burner If necessary sec...

Страница 5: ...the burner once it is completely cool Before cleaning close the gas shut off valve at the laboratory workbench gas isolation valve Do not use chemicals for cleaning Never drill or unscrew the burner n...

Страница 6: ...l not ignite despite open shut pff valve at the laboratory workbench gas isolation valve no audible gas flow The main gas supply has not been switched on gas bottle is empty The actuation knob for aut...

Страница 7: ...may only be changed by qualified personnel according to the DVGW or valid natio nal regulations 10 Guarantee All burners are checked before distribution If you discover any functioning fault please se...

Отзывы: