MT2 Trimmer

Nr. 1801-0000

ENGLISH

1. Introduction

We are pleased with your decision to purchase the

MT2

- Trimmer which sets new standards with respect

to function, performance, safety, and styling.

DSS Disc Stop System:

This trimmer is equipped with a maintenance-free
electronic motor brake. After the unit is switched off or
if the front cover is opened, the trimmer disc is stopped
within 5 seconds. The resulting buzzing sound
emanating from the motor area is perfectly normal.

Please read the following operating
instructions carefully and observe all the
information they contain in order to
ensure long and trouble-free service.

2. Application Area

The

MT2

is a wet trimmer used for grinding plaster or

acrylic models for dental technology applications.

2.1

Ambient Conditions

(in accordance with DIN EN 61010-1)

The unit may only be operated:

• indoors;
• up to an altitude of 2,000 m [6,500 ft] above sea

level;

• at an ambient temperature range of 5 - 40ºC [41 -

104ºF] *);

• at a maximum relative humidity of 80% at 31ºC

[87.8ºF], dropping linearly to 50% relative humidity
at 40ºC [104ºF] *);

• with a mains power supply whose voltage

fluctuations do not exceed 10% of the rated value;

• under contamination level 2 conditions;
• under over-voltage category II conditions.

*) Between 5 - 30ºC [41 - 86ºF] the unit can be operated at a

relative humidity of up to 80%. At temperatures between 31 -
40ºC [87.8 - 104ºF] the humidity must decrease proportionally in
order to ensure proper operability (e.g., at 35ºC [95ºF] = 65%
relative humidity, at 40ºC [104ºF] = 50% relative humidity). The
unit may not be operated at temperatures above 40ºC [104ºF].

3. Hazard Information

3.1

Employed Symbols

The following symbols are used in these instructions
and on the unit:

Danger
This indicates an imminent risk of injury.

Electrical hazard
There is a risk of electrical shock.

Attention
Failure to observe the associated
information can result in damage to the
unit.

Note
The operating instructions contain useful
information to make handling easier.

Wear eye protection.

For indoor use only.

Before opening the unit, disconnect it
from the mains power supply by pulling
the wall plug.

Note the operating instructions.

3.2

Hazard Information

During trimming, please observe the accident
prevention regulations of the relevant professional
association as well as the following safety information:

• The 

MT2

is an electrical device with inherent

hazard potential. This device may only be
installed and operated by authorized
specialists after examination of compliance
with locally applicable standards.

• The unit may only be connected to a properly

grounded power supply.

• Prior to commissioning, compare the voltage

information on the nameplate with the
specifications of the local mains power
network.

• Never touch the trimmer disc while the unit is

connected to the mains power supply. Serious
injury could result should the unit accidentally
start!

• Never wear loose clothing when working with

Содержание Klettfix MT2

Страница 1: ...iet 78247 Hilzingen Germany Tel 49 7731 8208 0 Fax 49 7731 8208 70 info renfert com www renfert com MadeinGermany Bedienungsanleitung Instruction manual Mode d emploi Istruzioni d uso Instrucciones para el servicio Gebruiksaanwijzing Инструкция по эксплуатации ...

Страница 2: ...C 87 8 104ºF the humidity must decrease proportionally in order to ensure proper operability e g at 35ºC 95ºF 65 relative humidity at 40ºC 104ºF 50 relative humidity The unit may not be operated at temperatures above 40ºC 104ºF 3 Hazard Information 3 1 Employed Symbols The following symbols are used in these instructions and on the unit Danger This indicates an imminent risk of injury Electrical h...

Страница 3: ...es Figure 3 Using the Allen key turn the centring screw clockwise to tighten it Figure 4 Turn the disc by hand to make sure it is running true Figure 5 If the disc exhibits large variations in height remove it turn it 180 reinstall Insert the M5 covering cap from the mounting hardware set into the Allen screw to protect the screw against contamination 4 Mount the spray pipe on the water inlet Figu...

Страница 4: ...d refer to the following section 7 Install the new trimmer disc Figure 3 8 Make sure the locking pins are properly seated in the retention holes Figure 3 9 Using the Allen key turn the centring screw clockwise to tighten it Figure 4 10 Turn the disc by hand to make sure it is running true Figure 5 If the disc exhibits large variations in height remove it turn it 180 reinstall it 11 Install the cov...

Страница 5: ... dealer must be provided when making warranty claims Components subject to natural wear e g trimmer discs are excluded from this warranty The warranty is voided in case of improper use failure to observe the operating cleaning maintenance and connection instructions in case of independent repairs or repairs by unauthorized personnel if spare parts from other manufacturers are employed or in case o...

Страница 6: ...ed The spraypipe is blocked Open the water supply Increase the water supply to the trimmer Open the water supply Clean de scale the pipe Water sprays out of the trimmer opening Too much water The water drainage is not functioning properly Reduce the water supply to the trimmer Check if the drainage pipe is attached properly gradient The disc is wobbling Disc is not lying flat against the joint Dis...