ELPIS-4 / 3 / 2 User Guide
2017.08.21 Ver M-1.0.0
37/45
10.
EMC emissions and immunity
Guidance and manufacturer’s Declaration – Electromagnetic Emissions
The ELPIS-4, 3, 2 System is intended for use in the electromagnetic environment specified below.
The customer or the user of the equipment should assure that it is used in such an environment
Emission Test
Compliance
Electromagnetic environment
– guidance`
RF emissions
EN55011
Group 1
The ELPIS-4, 3, 2 System is intentionally generated
and / or used conductively coupled radio-frequency
energy which is necessary for the internal
functioning of the equipment itself.
RF emissions
EN55011
Class A
The ELPIS-4, 3, 2 System is suitable for use in all
establishments other than domestic and those
directly connected to the public low voltage power
supply network that supplies buildings used for
domestic purposes.
Harmonic
Emissions
IEC 61000-3-2
Class A
Voltage
Fluctuations / Flicker
Emissions
IEC 61000-3-3
Complies
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The ELPIS-4, 3, 2 System is intended for use in the electromagnetic environment specified below.
The customer or the user of the equipment should assure that it is used in such an environment.
Immunity Test
IEC 60601
IEC 60601
Test Level
Compliance Level
Electromagnetic
environment - guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6 Contact
± 8 Air
± 6 Contact
± 8 Air
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the relative
humidity should be at least
30%
Electrical fast
Transient / burst
IEC 61000-4-4
± 2
㎸
for power
supply lines
± 2
㎸
for power
supply lines
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
± 1
㎸
differential
mode
± 2
㎸
common
mode
± 1
㎸
differential
mode
± 2
㎸
common
mode
Anti-surge protection needs to
be incorporated into the main
supply to the equipment if
surge protection is to be
guaranteed.