14
3.
ADVANCED CONFIGURATION (for technicians only)
The codes that are found in the unit are:
02000
Reception of the programme
and the transmitter’s version: switch on the transmitter
after switching on the instrument.
Continue with
TARE.
Now, a value that corresponds to the intensity of the reception field will be
displayed.
Confirm with
MODE.
00531
Serial configuration
Choose between enabling the serial output:(
S SI
) or inhibiting it (
S no)
with
MODE.
Continue with
TARE.
Select the transmission rate with
MODE (1200, 2400, 4800, 9600, 19200, 38400).
Note: if used in receiving mode, the baudrate is 2400
Continue with
TARE.
Choose the transmission type: continuous (
Conti
), with repeater (
riPEt
), with printer (
StAMP
) with
MODE.
Continue with
TARE.
If you chose continuous transmission, you can enable the transmission of the measurement unit (
U
SI
–
U NO
) by pressing
MODE.
If you chose transmission with repeater, you can choose the number of digits to be transmitted (
C 5
or
C 6
) by pressing
MODE.
If you chose transmission with printer, you can choose the printer type: 80 columns (
S 80c)
, 24
columns receipt (
S 24c
), 24 columns on one sole line (
S 24L
)
.
00400
Calibration code
Enter the code and press
MODE for a long time
to continue with the setting, Choose
the division value with MODE and TARE, confirm by pressing
MODE for a long time
.
Perform the zero setting by pressing
MODE.
Load a weight and press
TARE
, set the
loaded weight value and press
MODE for a long time
. If necessary, also calibrate the
following steps.
00022
Various configurations
Enter the code and press
MODE
to continue with the setting, the various parameters
are:
P Min
(
n
) This parameter is the weight value within which the weighing is considered to be minimum.
The measurement unit is the same as that of the displayed weight.
T 0tr
(
t
) This parameter defines a time range which, once over,
causes any zero variations to be compensated.
R 0tr
(
r
)
This parameter defines a weight range within which all zero
variations need to be compensated.
P MAS
(
P
)
(If envisaged), it defines a weight value which, once
exceeded, causes the display unit to lock out the reading of
weight and display
T PS
(
S
)
.This parameter defines a time range which, once elapsed,
causes a weight that is not moving to be considered as
stable.
R PS
(
c
)
This parameter defines a weight range within which a weight
is considered to be stable