53
Message
What it means
What to do
A used test strip was
inserted.
Remove the test strip
and insert a new test
strip.
The blood or control
solution sample was
applied before the
symbol appeared.
Repeat the test with a
new test strip and wait
until the symbol
appears before
applying the blood or
control solution sample.
The thermometer
of meter is not
stabilized yet.
Retry the test after
about 30 minutes for
the thermometer to be
stabilized.
The blood sample is of
insufficient volume or
blood is too thick.
Repeat the test using a
new test strip.
This error message may
appear when the
wrong blood glucose
test strip is used instead
of the ReliOn™ Premier
Test Strip.
Repeat the test with a
ReliOn™ Premier Test
Strip. Insert the test strip
with the contact bars
facing up, and push in
gently until the meter
turns on.
Understanding Error Messages