 9 

  Installation 

The equipment delivered is ready for immediate use. The sealed plug can be connected 

directly to the mains supply 230 V/16 A. For installation please refer to chapter 2. 

  Operation 

  Chronological run of operation 

1.  Operate the 

device only when it’s filled with water. (Pay attention to safety notes in 

chapter 2) In order to avoid an unwanted dry-run, a daily filling in the morning is 

recommended. Turn the heating controller (5) completely to the left (if necessary ventilate 

tank by activating steam gun). Note: Fill the device only in a cold state. Open the tank 

cap (6) and fill in distilled or decalcified water up to the edge (a filling funnel is 

recommended). Close the tank cap (6) and hand-tighten it. 

2.  Turn the device on with the switch POWER (2). Set heating control (5) to max. If you 

don’t set heating controller to max., the “low water level“ indicator lamp lights up (4) 

despite a filled tank. However the device is ready for operation, but at a low temperature. 

The heating starts automatically. After approx.18 min., pressure has been achieved and 

the indicator lamp goues out if the heating controller was is set to max.. 

3.  Start the hose heating with the switch HEATING (3) ein. 

4.  The maximum steam pressure can be modified with the heating controller (5). (Pressure 

is only built up in the upper quadrant of the controller range. Pressure is built up with 

delay.) 

5.  Open the steam valve to take out dry steam (1). 

6.  Device needs refilling: indicator lamp (4) lights up. For refilling please adhere to chapter 4 

Notes for the User, item 1. If the control lamp does not turn off automatically after filling, 

please press the temperature limiter (11). 

 

 

Содержание STEAMY PR

Страница 1: ...erating Instructions LABORATORY EQUIPMENT IN STAINLESS STEEL STEAMY PR REITEL Feinwerktechnik GmbH Senfdamm 20 49152 Bad Essen Phone 49 0 5472 9432 0 Fax 49 0 5472 9432 40 www reitel com info reitel c...

Страница 2: ...the device please check carefully whether transportation damages of any kind have occurred If that is the case the carrier or manufacturer must be notified within 3 days The company will not be respon...

Страница 3: ...TION 6 VERSION WITH FLAT NOZZLE 6 TECHNICAL DATA 6 SAFETY NOTES 7 GENERAL SAFETY NOTES 7 SPECIFIC SAFETY NOTES 8 INSTALLATION 9 OPERATION 9 CHRONOLOGICAL RUN OF OPERATION 9 DECALCIFYING THE DEVICE WHE...

Страница 4: ...Order No 10205000 Water softening device DELIMER COMFORT Operating Elements 1 Steam valve 2 POWER switch 3 Swith to activate hose heating HEATING 4 Indicator lamp low water level 5 Heating control 6 T...

Страница 5: ...5 Device view Backside 1 13 11 5 4 6 7 3 9 10 2 12 14 8...

Страница 6: ...lat nozzle Technical Data Line voltage 230 V 50 60 Hz Current consumption 2 300 W Height width depth 420 300 350 mm Weight 14 kg Pressure 4 bar Temperature 150 C Volume 5 l Steam level 1 Filling manua...

Страница 7: ...e the manufacturer s warranty expires The device should be set up in a clean well ventilated room if necessary under a hood Supporting surfaces should be straight flat and adequately stable Don t plac...

Страница 8: ...cale deposits is not covered by warranty Decalcification is depending on water hardness every one or two weeks Verwenden Sie hierf r handels bliche Entkalkungsmittel und beachten Sie die Herstellerang...

Страница 9: ...the device on with the switch POWER 2 Set heating control 5 to max If you don t set heating controller to max the low water level indicator lamp lights up 4 despite a filled tank However the device is...

Страница 10: ...vice cool itself down and empty the tank 3 Fill the tank with decalcifier according to manufacturer s instructions 4 Close tank cap 6 5 Turn switch POWER 2 on As soon as the pressure starts building u...

Страница 11: ...or seal No steam a Cold water was refilled and device is heating up b Device is calcified blocked valve switches or valves defective Call customer services No steam pressure 1 Regulator 5 was not set...

Страница 12: ...g the device when not using distilled water Clean the housing regularly with a moist cloth Note Clean and protect your device regularly with the REITEL stainless steel care spray Order No 90405000 to...

Страница 13: ...ssed to the distributor the product was purchased from or directly to our service department Contact REITEL Feinwerktechnik GmbH Senfdamm 20 49152 Bad Essen Germany Phone 49 0 5472 9432 32 Service Dep...

Страница 14: ...ages due to faulty installation or insufficient maintenance of the product e g decalcification Damages due to natural disaster e g lightning water fire act of nature beyond control war inadequate line...

Страница 15: ...dvised to backup essential programs and data before sending for service 12 This warranty shall be interpreted in accordance with German law to the exclusion of UN sales law Place of jurisdiction is ou...

Страница 16: ...16 Disposal Regulations According to valid laws our devices are disposed of according to B2B Regulations for proper disposal Please contact your local dealer for additional information...

Страница 17: ...ng Rubber ring 1 St pc E6227700 Heizk rper inclusive Flansch und Dichtung Heating element incl flange and seal 1 St pc E7327600 Isoliermatte nicht gezeichnet Isolation mat not drawn 1 St pc E6021200 K...

Страница 18: ...r Schlauchheizung 24 V 75 W Transformer for hose heating 24 V 75 W 1 St pc E6175500 Verschraubung UNI VS 13 ohne DK 030 Screw joint UNI VS 13 without DK 030 1 St pc E7090000 Verschraubungen bitte For...

Страница 19: ...19 EU DIRECTIVE...

Страница 20: ...pliance with Quality Management ISO 13485 2016 incorporating established standards REITEL Feinwerktechnik GmbH Senfdamm 20 49152 Bad Essen Phone 49 0 5472 9432 0 Fax 49 0 5472 9432 40 www reitel com i...