Dimensions -
54 x 27 x 100 mm / 2.13 x 1.06 x 3.9 inches
Power -
2 AAA Batteries, included.
Battery life -
10,000 readings, Auto Off Sleep Mode.
Ratings -
IP65 Dustproof/Water Resistant, CE, RoHS, and WEEE
compliant.
Reading Modes -
Brix-TC = Percent sucrose in aqueous solution at
20 Degrees C.
RI-TC = Refractive index temperature corrected
to 20 degrees C.
RI = Refractive index
Warranty -
One year against manufacturing defects. Evidence of
tampering voids warranty.
Reichert Analytical Instruments
3362 Walden Ave
Depew, NY 14043
USA
Toll Free: 888-849-8955
Phone: 716-686-4500
email: [email protected]
www.reichertai.com
Reichert GmbH
Hubertusstrasse 2
D-82229 Seefeld
Germany
Tel: +49-8152-993530
ISO-9001 Certified
13940000-101 REV E
SPECIFICATIONS:
Catalog
13940000
RI Range -
1.3330 - 1.4465, Accuracy +/- 0.0002
BRIX Range -
0 - 62%, Accuracy +/- 0.2%
Auto Temperature Compensation -
20
o
C RI and BRIX
Calibration -
Distilled Water
Prism -
Glass
Illumination -
589nm LED
ERROR CODES:
Err01 - No sample present. Add sample.
Err02 - Inadequate sample. Add additional sample.
Err03 - Sample exceeds the refractive index reading range.
Err04 - Sample temperature has not equilibrated. Allow more time for
temperature equilibration.
Err05 - Excessive ambient light. Cover sample well when reading.
Err06 - Excessive ambient light. Cover sample well when reading.
Err07 - Positive calibration error. Re-calibrate with distilled water.
Err08 - Negative calibration error. Re-calibrate with distilled water.
Err09 - Poor sample condition. Sample may not be able to be read.
Err10 - Sample type may not be able to be read.
Err12 - Index of sample is out of range of instrument.
Err5X - For any errors in this series contact Reichert Analytical
Instruments for technical assistance.
BATTERY CONDITION:
The following display examples represent either a low battery condi-
tion or an extremely low battery condition where the instrument will
no longer operate properly. In either case the batteries should be
promptly replaced.
Battery condition low.
Replace Batteries.
Battery condition extremely low.
Replace Batteries.
TEMPERATURE COEFFICIENT CALCULATIONS:
For non-sucrose based samples, should you need to temperature
correct the value, visit our web site at www.reichertai.com. Please
download a copy of “Temperature Coefficient Calculation and Tem
-
perature Correction” found under the “APPLICATION” section of our
site.