Reichert 13940000 Скачать руководство пользователя страница 2

Dimensions -

 54 x 27 x 100 mm / 2.13 x 1.06 x 3.9 inches

Power - 

2 AAA Batteries, included.

Battery life - 

10,000 readings, Auto Off Sleep Mode.

Ratings - 

IP65 Dustproof/Water Resistant, CE, RoHS, and WEEE  

  compliant.

Reading Modes - 

Brix-TC = Percent sucrose in aqueous solution at  

                 20 Degrees C.

RI-TC = Refractive index temperature corrected  

 to 20 degrees C.

 RI   =     Refractive index

Warranty - 

One year against manufacturing defects. Evidence of  

     tampering voids warranty.

Reichert Analytical Instruments

3362 Walden Ave

Depew, NY 14043

USA

Toll Free: 888-849-8955

Phone: 716-686-4500

email: [email protected]

www.reichertai.com

Reichert GmbH

Hubertusstrasse 2

D-82229 Seefeld

Germany 

Tel: +49-8152-993530

ISO-9001 Certified

        13940000-101 REV E

SPECIFICATIONS:

Catalog 

13940000

RI Range - 

1.3330 - 1.4465,   Accuracy  +/- 0.0002

BRIX Range - 

0 - 62%,   Accuracy  +/- 0.2%

Auto Temperature Compensation -

 20

C RI and BRIX

Calibration -

 Distilled Water

Prism -

 Glass

Illumination -

 589nm LED

ERROR CODES:

Err01 - No sample present. Add sample.

Err02 - Inadequate sample. Add additional sample.

Err03 - Sample exceeds the refractive index reading range.

Err04 - Sample temperature has not equilibrated. Allow more time for  

            temperature equilibration.

Err05 - Excessive ambient light. Cover sample well when reading.

Err06 - Excessive ambient light. Cover sample well when reading.

Err07 - Positive calibration error. Re-calibrate with distilled water.

Err08 - Negative calibration error. Re-calibrate with distilled water.

Err09 - Poor sample condition. Sample may not be able to be read.

Err10 - Sample type may not be able to be read.

Err12 - Index of sample is out of range of instrument.

Err5X - For any errors in this series contact Reichert Analytical 

            Instruments for technical assistance.

BATTERY CONDITION:

The following display examples represent either a low battery condi-

tion or an extremely low battery condition where the instrument will 

no longer operate properly. In either case the batteries should be 

promptly replaced.

      Battery condition low. 

      Replace Batteries.

      

      Battery condition extremely low. 

      Replace Batteries.

TEMPERATURE COEFFICIENT CALCULATIONS:

For non-sucrose based samples, should you need to temperature 

correct the value, visit our web site at www.reichertai.com. Please 

download a copy of “Temperature Coefficient Calculation and Tem

-

perature Correction” found under the “APPLICATION” section of our 

site.