3 INSTALLATION AND CONNECTION 

 

3.1 PRE-INSTALLATION CHECK   
Remove the parts of the device from the box and place it in a safe, stable position. 
Check the accessories according to the packing list to ensure integrity.

3.2 CHARGING METHOD 
When the Laser Hero is charging, the power
indicator is yellow; when fully charged, the power
indicator turns blue. The charging time reaches
85% for about 6 hours.

 

NOTICE:

 

•  For first time use or if the device has not been used for a long time, it is recommended

 

  to fully charge the battery before use.

 

•  Prevent bumps and drops when the instrument is held or placed 

 

•  Use only our company approved power adapters and batteries that meet the safety

 

  standards of IEC60601-1. Unauthorized power adapters can cause the device battery to explode 

 

  or damage the device. 

 

•  The device can be used while charging; This may extend the battery’s full charge time. 

 

•  If the power supply voltage is unstable during charging, and the light source cannot be started 

 

  normally, the charger can be unplugged from the device. 

 

•  The device generates a certain amount of heat when it is charged, which is a normal phenomenon 

 

  and does not affect the performance and service life of the device. However, if the battery is 

 

  abnormally overheated, stop charging immediately. 

 

•  If the battery does not charge properly, please contact our customer service department.

 

•  Disconnect the device from the charger after fully charging. Unplug the charger from the device 

 

  and unplug it from the power outlet. 

 

•  To save power, unplug the power adapter when not in use. When charging, the charger should be 

 

  in full contact with the outlet. 

 

•  The instrument should be installed and operated to ensure normal working environment, power 

 

  supply, and electromagnetic compatibility requirements. Refer to the data provided in

 

  Appendix F to ensure a normal working environment. 

 3.3 INSTALLING THE INNER LINER  
Paste the inner liner of the appropriate
thickness according to the size of the head.

4 OPERATION OF INSTRUMENT 

 

4.1 PREPARATION BEFORE TREATMENT 
Before treatment, clean the scalp and dry the hair. .

4.2 OPERATION OF LASER HERO 

4.2.1 Instrument operation method 

1)   Effective irradiation zone
    of the light source is shown
    in the FIGURE 4-1.
2)  According to the patient’s head
    circumference, a suitable inner liner is
    pasted on the inner side of the hair loss
    therapy helmet (adjustable sizing).
3)  Press and hold the “      ” button until lighting and the working mode prompt voice will
  say 

“Welcome”.

4)  Press the “       ” button to select the treatment time (25 min, 30 min, 35 min, unit: minutes).
    If not, the default treatment time is chosen. Press the “       ” button to flash and the voice prompt 
    “Treatment begins, please wear the helmet” will be sent out. Once the helmet is placed on your 
    head, the laser is turned on and the treatment begins. The prompt voice “The remaining treatment 
    time is 5 minutes” will sent out when there is 5 minutes treatment time remaining. When the 
    treatment ends, the prompt voice “Treatment has ended,thank you for using” will be sent out.
5)  It is recommended to irradiate every two days.
6)  In the course of treatment, if pressing the “      ” button again, the treatment will be stopped
    and the “      ” button will stop flashing.
7)   In the course of treatment, if the hair loss therapy helmet is removed, the irradiation will be
    paused and the prompt voice “Treatment stopped,the helmet will be turned off in 3 minutes”
    will be sent out. If the hair loss therapy helmet is re-worn within 3 minutes, the irradiation will 
    continue.with a prompt voice“Continue for treatment”.If the hair loss therapy helmet is not put
    on within 3 minutes, the instrument will be automatically shut down.
8)   Press and hold the “       ” button, voice on or off can be chosen. When the “      ” button is 
    flickering, the instrument is under the waiting for treatment or working state. At this time,the “      ” 
    button cannot be operated.
9)   Press and hold the “      ” button, indicator lights shut off, and power ends.

 

 

08  |  09

Содержание LASER HERO

Страница 1: ...U S E R M A N U A L...

Страница 2: ...ult in damage to the device or a person Our company does not assume the related liability if there is device damage or personal injury caused by not following the instructions specified in this user m...

Страница 3: ...ed Please keep this manual near your Laser Hero so that it can easily be obtained when needed 1 3 Home Use The system should be used under the guidance of a doctor or specialist Contact your doctor im...

Страница 4: ...hours KN 8000C Working environment TEM 5 30 C Relative humidity 85 Atmospheric pressure 700hPa 1060hPa Structure helmet type Display mode LED display Effective irradiation area 370cm2 10 Laser classi...

Страница 5: ...ent 3 3 INSTALLING THE INNER LINER Paste the inner liner of the appropriate thickness according to the size of the head 4 OPERATION OF INSTRUMENT 4 1 PREPARATION BEFORE TREATMENT Before treatment clea...

Страница 6: ...commended that the user perform a comprehensive technical inspection of the instrument and accessories every six months to check for mechanical damage and cable damage 4 If the service life of the ins...

Страница 7: ...ies Do not pierce the battery with nails or other sharp objects The battery should be protected from exposure or direct sunlight avoid extreme radiation infrared and ultraviolet radiation and avoid co...

Страница 8: ...LASER HERO USER MANUAL 14 15...

Страница 9: ...1760 Wadsworth Rd Akron OH 44320 1 800 239 7880 Fax 1 330 564 0118 info paintechnology com paintechnology com...

Отзывы: